3 • Operation and Maintenance
Doc. 065051-03 1/08
85
Example Program for Concentrate Mode (Matrix Elimination)
The following example Chromeleon or Chromeleon Xpress program
shows the AS commands when the autosampler is in concentrate mode.
Commands similar to these could be used for a matrix elimination
application.
Flush
Volume = 100
;AS options commands
Wait
FlushState
NeedleHeight =
2
CutSegmentVolume =
0
SyringeSpeed =
3
CycleTime =
0
Pipet Volume = 20.0, SourceVial =
CurrentVial, DestinationVial =
CurrentVial
;AS sample prep
commands
Mix SourceVial = CurrentVial,
NumberOfTimes = 5, Volume = 250.0
FlushSP Volume = 250.0
DelaySP Time = 0.0
Dispense Volume = 1000.0,
SourceReservoir = Reservoir_A,
DestinationVial = CurrentVial
WaitForTemperature = False
Concentrate ValvePosition =
LoadPosition, AspirateSpeed = 3,
DispenseSpeed = 1
;Switch the inject valve
to Load and load sample
onto the concentrator
ReagentPrime Volume = 10000.0,
SourceVial = Reservoir_A, ValvePosition
= NoChange
;Prime sampling needle
assembly with reagent
ReagentFlush Volume = 5000.0, SourceVial
= Reservoir_A, ValvePosition = NoChange
;Flush concentrator with
reagent; this is the
matrix elimination step
Wait
SampleReady
0.000
Wait
CycleTimeState
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...