AS Autosampler Operator’s Manual
306
Doc. 065051-03 1/08
D.8 Aligning the Sampling Needle from the AS Front Panel
Two methods are available for aligning the sampling needle:
•
Automatic alignment, using Chromeleon or Chromeleon Xpress (Release 6.8
SP2 or later) (recommended method) (see
•
Manual alignment, using the front panel screens
This section describes the procedure for aligning the sampling needle manually
using the front panel screens. Dionex recommends manual alignment only if
Chromeleon or Chromeleon Xpress (Release 6.8 SP2 or later) is unavailable.
Align the sampling needle in the inject port after replacing the sampling needle
assembly and/or the needle seal assembly, or if the needle becomes misaligned at
anytime.
NOTE Before aligning the sampling needle, always inspect the
needle to make sure it is not bent. If the needle is bent,
replace the sampling needle assembly (see
NOTE This procedure has numerous steps. To help maintain
your place in the procedure, check off each step after
completing it.
1. Close the autosampler door and turn on the power.
2. (Optional) For a better view of the needle and inject port during the alignment
procedure, temporarily disable the door lock to allow operation with the door
open.
a. On the
MENU of SCREENS
, press
5
and then
7
to go to the
DOOR
INTERLOCK BYPASS
screen.
Figure D-15. Door Interlock Bypass Screen
Help Prompt
DOOR INTERLOCK BYPASS
DOOR INTERLOCK BYPASS
NOTE: After setting bypass you have a
15 minute window to perform operations
with the sample door open.
YES
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...