3 • Operation and Maintenance
Doc. 065051-03 1/08
91
Setting Up a Sequential Concentrate Mode Injection
The AS must be equipped with the sequential injection option
(P/N 063294) in order to perform sequential injections. In general, the
following setup steps are required:
1. Follow the instructions in
to install the sequential
injection diverter valve, connect the AS inject port tubing to the
diverter valve, and connect the diverter valve to the injection valves
on two systems.
2. Install a concentrator column on each injection valve.
3. Calibrate the inject port volume of each system. This is the volume
from the inject port through the diverter valve to the injection valve
(see
4. Follow the instructions in
injection, except select
SEQUENTIAL CONCENTRATE
in the
SAMPLE
MODE
field on the
SYSTEM PARAMETERS
screen (see
Figure 3-30. System Parameters Screen: Sequential Concentrate Mode
Help Prompt
SYSTEM PARAMETERS
WAIT:
ENABLED
METHOD SYRINGE SPEED:
CUT SEGMENT VOLUME:
INJECT PORT VOLUME:
uL
uL
10
S1 43
3
SAMPLE MODE:
SEQ CONC
FLUSH SYRINGE SPEED:
4
uL
S2 47