AS-HV Operator’s Manual
8
Doc. 065125-01 10/06
2.2
AS-HV Side Panel
This section describes (from top to bottom) the components located on the right
side panel of the AS-HV.
ERROR Indicators
If a fault occurs, the
ERROR
indicators light to
display a binary error code (in blue) that identifies
the fault. For example, the error code
0110
indicates
that Error 6 (an X-axis position error) has occurred. Refer to
explanation of all of the error codes.
RESET Switch
Pressing the
RESET
switch returns the AS-HV to its
initial start-up sequence. The reset procedure identifies
problems with autosampler movement in the X-, Theta-, and Z-axes (see
). Successful completion of the reset procedure confirms that the
motors operate and the action of the carriages is correct.
A small pointed device is required to depress the recessed switch.
DIP-A DIP Switches
The DIP switches are numbered 1 to 8 (from top to bottom).
SW5 sets the baud rate (the only communication parameter
that can be set on the AS-HV).
The table below lists the DIP switch functions and default
settings. The switch positions are indicated as 0 = Off
(Up/Open) or 1= On (Down/Closed).
DIP Switch
Default Setting
Description
SW1
Off
Self-test mode
SW2
Off
Reserved
SW3
Off
Reserved
SW4
Off
Reserved
SW5
On
Baud rate
SW6
Off
Mode selection
SW7
Off
Mobile rinse station expiration timer
SW8
Off
Wash select MRS/FWR (mobile rinse
station/fixed rinse reservoir)
Содержание AS-HV
Страница 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Страница 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Страница 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Страница 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Страница 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Страница 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Страница 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...