DOSAGE

 

 

 

47 

 

If during the dosage the mains voltage is cut off, when it is restored, the instrument will turn on 

automatically into PAUSE; press the 

 button to continue with the interrupted dosage. The 

dosage now restarts from the condition in which it was in, when it turned off. 
 
In the PAUSE condition it is possible to access the USER MENU function "Repeat dosage cycles", in which it 
is possible to change the number of the cycles. 
It is possible to enter in Repeat dosage cycles step only in weights formula mode. 
It is also possible to turn off the instrument by pressing at length the C key. 
 

6.13.3

 

Dosage cycle reset 

To cancel the dosage cycle in execution, keep closed the pause input(IN.2) for about 3 seconds or push 

 button. The "RESET DOSAGE" message will be shown on the display. 

By carrying out this operation the indicator cancels the dosage and the GROSS weight on the scale plate 
appears on the display. By cancelling the dosage under way, if the material has been dosed, the production 
of the single formula and the general total is increased. 
 
To cancel a dosage during a repetition of the dosage cycles, one should carry out a reset (described 
previously); the status bar will show "RESET DOSAGE"; after this, the instrument is in the WAIT FOR START 
status: by giving a start command the number of executed dosages shown on the display is increased by 1, 
and the dosage continues with the following cycle. 
 
To cancel a set repetition of cycles, one should execute a reset in order to exit the dosage cycle: the 
message “RESET DOSAGE” will appear for a few instants on the display and the instrument goes into the 
WAIT FOR START status. At this point one should execute a further reset, which will cancel the set 
repetitions.The number of repetitions is zeroed, returning into the 1 status. 
NOTE: the reset does not work in the WAIT FOR START condition (except for when the number of 
repetitions is cancelled). 
 

6.13.4

 

Setting the number of consecutive repetitions of the active formula 

It is possible to programme a series of automatic repetitions of the dosage cycles using a specific formula. 
 
One should: 

Push 

 button 

Enter the number of repetitions to be executed (value which is between 0 and 999). 
 

Formula Cycles = 0 

 infinite repetitions with manual start through the START button 

 

Formula Cycles = 999 

 infinite repetitions with automatic start 

 

Formula Cycles = n 

 executes n repetitions with automatic start  

Confirm the number of set cycles by pressing the OK key. 
 
DESCRIPTION OF THE FUNCTION: 
The display will show the number of executed cycle repetitions, which will increase at the end of each 
complete cycle, and the number of cycles to be executed. 

Содержание EGT scale series

Страница 1: ...EGT BATCH USER GUIDE SOFTWARE FOR MULTICOMPONENT DOSAGE SYSTEMS Revision 1 00 Last update 08 05 2013...

Страница 2: ...Page intentionally left blank...

Страница 3: ...lphanumeric input 19 4 2 Keyboard 19 4 3 Cross light 19 5 MAIN FUNCTIONING DESCRIPTION 21 5 1 Zeroing 21 5 2 Tare 22 5 2 1 Semiautomatic tare 22 5 2 2 Preset tare 23 5 2 3 Link a preset tare to a prod...

Страница 4: ...timer 42 6 12 Formula execution modes 43 6 12 1 Formula weighs 43 6 12 2 Setting of the total weight 43 6 12 3 First free component 44 6 13 Dosage procedure 45 6 13 1 Dosage cycle starts 46 6 13 2 Pau...

Страница 5: ...ing of the remote components dosage 50 6 18 Example of the formula and the description of the executed cycle 52 6 19 Error messages 55 6 20 Progressives 57 6 20 1 Additional value 57 6 20 2 Progressiv...

Страница 6: ......

Страница 7: ...rol a printer and communicate with a personal computer or to be inserted in a network of weight indicators controlled by a PC Please note that this instrument is covered by a warranty and MUST NOT BE...

Страница 8: ...e negative effects on the environment and health and supports the recycling of the materials of which the equipment is made The unlawful disposal of the product by the user will entail fines foreseen...

Страница 9: ...ms Texts database of 15 items SIGNALS 3 way control light KEYBOARD Impermeable polycarbonate keyboard IP65 protection degree with membrane keys with audible and tactile feedback PARAMETER SETUP Calibr...

Страница 10: ......

Страница 11: ...e Figure 1 It can be simply put on a table or fixed to a shelf or column available on request NOTE If the identification plate is supplied separately therefore not attached to the indicator it is advi...

Страница 12: ...ter which converts the 100 240Vac 50 60Hz mains voltage To power the instrument through the 240 Vac mains insert the plug and the adapter to the 240 Vac mains socket To connect the indicator to the po...

Страница 13: ...nts EGT AF05 XX is the name of the installed software in which XX identifies the language XX YY is the installed software version 3 The zeroing procedure clears the weight on the scale within the tole...

Страница 14: ......

Страница 15: ...15 4 INDICATOR PARTS Figure 3 Front panel Part Description 1 Touch screen 2 Cross light 3 colors 3 Keyboard 2 3 1...

Страница 16: ...CATOR PARTS 16 4 1 Display The indicator presents 2 main screens illustrated in the below scheme Figure 4 First main screen Figure 5 Second main screen 4 2 3 6 1 5 7 8 11 1 10 12 13 14 15 16 18 19 17...

Страница 17: ...started it displays the working activity name MISSING DOSED The working formula necessary to be dosed weight when indicator is in dosage DOS CYCLE Dosage cycles necessary to be excuted if repetition c...

Страница 18: ...weighing system is near the zero included within the interval of 1 4 and 1 4 of the scale division The weight is unstable A tare value has been acquired A preset tare value has been entered Active we...

Страница 19: ...left or right Clear clears all the text Bksp backspace 2 F switches to special characters Shift changes the character case and switches between letter and number modes OK exit saving the text Esc exi...

Страница 20: ...is enabled The operator starts dosing in an approximate mode in other words at maximum speed Once the weight is greater than TARGET TOLERANCE and lower than TARGET the green is enabled while yellow l...

Страница 21: ...tout Print functions management Generic functions Generic operations to the indicator lock keyboard calculator Diagnostic Functions to check the peripheral units working state Input texts Input texts...

Страница 22: ...or turns on 5 2 Tare 5 2 1 Semiautomatic tare Step Description Screen 1 Press the key or touch on the weight area to tare the gross weight on the scale 2 A message appears on the display 3 If the weig...

Страница 23: ...By selecting the product the preset tare is activated 5 2 4 Tare cancellation It s possible to cancel the stored tare value in the following ways By touching the weight area or by pressing the when t...

Страница 24: ...access to the input texts 2 Touch on the desired text to be changed 3 Insert the content of the text and confirm with the OK button Touch on the ESC button to go back to the main screen 5 4 Databases...

Страница 25: ...element in the first empty position 3 Touch the first field to insert the content 4 Insert the content of the field and confirm it with the OK button 5 Repeat the operation for the desired fields and...

Страница 26: ...ccess to the desired database 2 Touch the element or the A Z button to search the element to be updated 3 Touch the desired field to be modified 4 Insert the content of the field and confirm it with t...

Страница 27: ...Screen 1 Touch the button to get access to the desired database 2 Touch the element or the A Z button to search the element to be deleted 3 Touch the DELETE button 4 Press the button to confirm or no...

Страница 28: ...ed database 2 Touch the desired element to select 3 The description of selected element appears on the display 5 4 5 Deselection Step Description Screen 1 Touch the button in the first main screen to...

Страница 29: ...vious selected element disappears on the display 5 4 6 Alphanumeric search Step Description Screen 1 Touch the A Z button in the database update selection screen 2 Insert the first characters of the e...

Страница 30: ...the net weight must be almost 20 divisions With the non approved instrument the net weight must be greater than zero 1 2 HEADER Print the header 2 3 START CYCLE Press button Start cycle printing 3 4 S...

Страница 31: ...ET CYCLE Reset cycle printing 22 23 DOSAGE ERROR Dosage error printing 23 24 GENERAL TOTAL Press button General total printing 24 25 START TIMER Start timer printing 25 26 START PAUSE Start pause prin...

Страница 32: ...ties are AUTOMATIC DOSAGE MANUAL DOSAGE UNLOADING TIMER PAUSE MANUAL WEIGHT REMOTE DOSAGE SYNCHRONIZATION Each activity executes a specific function shown in the following sections It is possible to a...

Страница 33: ...lls on the scale and the TARGET is reached Once the wait time is finished i If the instrument is enabled the checks start on the dosage result if the weight of the dosed material is not within the set...

Страница 34: ...rator can remove the exceeding weight or add the missing weight in this phase it is possible to use the message which appears MANUAL DOSING OK if within the tolerance or MANUAL DOSING OUT TOLERANCE if...

Страница 35: ...t value is reached in which a SLOW WEIGHT is missing from the FINAL WEIGHT 2 At this point the MANUAL DOSING OUT TOLERANCE is displayed indicating the end of the fast dosage and the beginning of the s...

Страница 36: ...C loading of the silo D unloading of the silo 6 4 2 Split total unloading a Description This activity has the purpose of completely unloading the scale after one or more loading dosages SPLITTING the...

Страница 37: ...ding activity see previous unloading method for details If the WAIT FOR DISCHARGE END is set only the total unloading will end and the relative outputs be disabled as soon as the weight reaches the TA...

Страница 38: ...NG DOWN in which one presumes that material still falls on the scale and the TARGET is reached Once the wait time is finished If the instrument is enabled the check on the dosed weight starts if the w...

Страница 39: ...erials necessary for this activity must be manually stopped During the PAUSE activity it is possible to pause the dosage through the external INPUT 2 input or the PAUSE button disabling all the active...

Страница 40: ...e printing or the storage on PC PLC A wait time starts equal to the DATA READING TIME in order to allow the reading of the activity data the instrument waits for this time before passing to the follow...

Страница 41: ...T button so that 1 The eventual linked output will be disabled 2 The data of the executed activity is automatically stored and transmitted through the serial line for the printing or the storage on PC...

Страница 42: ...To end the activity If the 485 address is set at 255 when all the remote dosages are terminated If the 485 address is set at a specific address for a remote dosage when that dosage is terminated A wa...

Страница 43: ...asis of this total the instrument recalculates the weights to be dosed of all the loading activities which make up the formula proportionally to the target of the basic formula The recalculation of th...

Страница 44: ...in the last cycle With various dosage cycles the instrument executes the set repetitions with the same quantity of material ENTERING THE TOTAL WEIGHT 1 Press the button 2 Enter the total weight to be...

Страница 45: ...ht must be within the range shown on the display The maximum enterable value is 999999999 EXAMPLE BASIC FORMULA RECALCULATED FORMULA ACTIVITY BASIC TARGET ACTIVITY RECALCULATED TARGET Manual dosage or...

Страница 46: ...nter the number of repetitions to be executed For further details see section Setting the number of consecutive ripetition of the active formula 6 13 1 Dosage cycle starts To start the dosage cycle pr...

Страница 47: ...ument is in the WAIT FOR START status by giving a start command the number of executed dosages shown on the display is increased by 1 and the dosage continues with the following cycle To cancel a set...

Страница 48: ...he weight During entering the formula a tolerance can be entered expressed as a weight percentage percentile value on the target weight entered in the formula therefore the tolerance weight can be cal...

Страница 49: ...ws to obtain a progressive reduction of the correction value and therefore a greater accuracy of the actual flight weight this action can be buffered more or less by changing the programmed percentage...

Страница 50: ...er of formula in execution 6 16 4 Production end terminate The present production will be terminated when the present production cycle is totally executed It will ignore the rest production cycles 6 1...

Страница 51: ...when the input 1 is closed and the dosage starts the display shows ERROR PC mode is not 485 message see the section 4 2 5 The 485 address number of the local indicator must be different with remote i...

Страница 52: ...instrument which is dosing c Only when the remote dosage 1 is terminated the remote dosage 2 starts to execute dosage d Only when the remote dosage 2 is terminated the automatic dosage starts 6 18 Exa...

Страница 53: ...WEIGHT FLIGHT WEIGHT SLOW WEIGHT 50 5 5 kg is reached the instrument disables OUT2 the fast dosage is over and the dosage starts at the minimum speed f Once the TARGET WEIGHT FLIGHT WEIGHT in other wo...

Страница 54: ...utput linked to ACTIVITY 3 and waits for 20 seconds o Disables OUT7 enables OUT2 and OUT8 which is the output linked to ACTIVITY 4 and commands the complete unloading When the weight is lower than the...

Страница 55: ...step The scale weight is too much if the formula starts the total loaded weight will be higher than the capacity Unloaded the scale manually The formula target of single activity is not correct or the...

Страница 56: ...ck the connection and the configuration of the DGTQF and EGTBATCH The PC com communication mode is not on 485 when remote dosage activity is trying to start To set the PC com communication mode workin...

Страница 57: ...ssive digits In MENU progressives Progressive digits it s possible to set the number of digits with which the TICKET PROGRESSIVE and LOT PROGRESSIVE are printed The number of settable digits varies fr...

Страница 58: ...the clearing of the AUTOMATIC LOT up to 6 digits before carrying out the actual clearing one is asked to confirm or cancel the operation The AUTOMATIC LOT is a progressive number which increments of...

Страница 59: ...used during transport If onsite service is requested or necessary for any reason where the instrument is used the Customer will pay for all of the service technician s costs travel time and expenses p...

Отзывы: