DF702_LoRa Configuration manual

V2.0

Date 2018-5-16

Содержание DF702

Страница 1: ...orp Ltd Tel 86 10 64718863 64753700 http www dingtek com E mail service dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 1 DF702_LoRa Configuration manual V2 0 Date 2018 5 16 ...

Страница 2: ...OT Leading Brand The purpose of Measurement Instrumentation and Sensors Handbook CNDingtek is to provide a reference that is both concise and useful for engineers in industry scientists designers managers research personnel and protocol as well as many others who have measurement problems We provide to our Customer Protocol Installation Operation and Configuration detailed of devices A Smart City ...

Страница 3: ...rs 8 5 1 Required tools for configuration of LoRa Device 8 5 2 Connect TTL to sensor 9 5 3 Environment to Configuration of Device 11 5 4 Video link 15 6 Configurations of Relevant Parameters of Device 15 6 1 Required tools 15 6 2 Connect TTL to sensor 16 6 3 Use serial software to configure 18 6 4 Video link 21 7 Upgrade devices 21 7 1 Instructions 21 7 2 Tools 21 7 3 Connect TTL to sensor 21 7 4 ...

Страница 4: ...ight value upload interval data etc Equipment upgrades For more information if you have any questions please contact us first 2 Device Function Descriptions Note Device will upload data when its power is activated it can send us data automatically to the user through LoRa Module Main Function of Device LoRa Module DF702 Sensor can give detailed about the using place for example Trash Bin its full ...

Страница 5: ...tional Descriptio n Default Alarm Threshold Whether the data will be reported immediately when the alarm condition is reached Remarks Full Check if the Trash bin is full 30cm Can be modified No The Device will detect the Distance between sensor and garbage if this Distance is more Then 30cm no alarm if less than 30 cm alarm will be activated Fire Detect if a fire has been occurred 75 Can be modifi...

Страница 6: ...aced 4 1 Connection of Hollow Sockets Power socket connect the battery Hollow positions Configure the LoRa module parameters or debug module the jumper cap should be placed in a hollow position Solid position The default position the device to transmit data configure device parameters such as data upload interval and full height the jumper cap should be placed at right side a solid position Band T...

Страница 7: ...ang District Beijing China 100022 Page 7 Figure4 1 4 2 Reverse side of picture of PCB board open the Screws of PCB Board as shown in figure 4 2 TXRX GND Hollow position Solid position Black part of the jumper cap the default in this position This interface can be used to modify Lora parameters as well as debug module This power socket is used to connect the Battery ...

Страница 8: ...evant parameters such as devaddr deveui and nwkskey are helpful in configuration process Required tools for configuration of LoRa Device TTL and its cable Serial software LoRa instruction manual Equipment TTL and cable for the goods accessories Please check these two items when you receive package from company as shown in Figure 5 1 and Figure5 2 GND TX RX This interface is used to configure devic...

Страница 9: ...rvice dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 9 Figure5 1 Figure5 2 Serial software 5 2 Connect TTL to sensor Device Serial Port connection GND TX and RX locations have been marked on the diagram as shown in Figure5 3 TTL TTL Cable ...

Страница 10: ...nbai Road 12 Chaoyang District Beijing China 100022 Page 10 Figure5 3 Connect TTL to device as shown in Figure 5 4 Figure5 4 Steps 1 Change jumper cap to hollow position default it is in the solid position as shown in figure5 5 2 Connect the TTL to the device as shown in figure 5 4 3 Connect the device s power supply as shown in figure5 6 GND RX TX ...

Страница 11: ...k com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 11 Figure5 5 Figure5 6 5 3 Environment to Configuration of Device 1 Open the serial software when opening the default information of serial port software is shown in figure 5 7 Jumper cap is in hollow position ...

Страница 12: ... 12 Chaoyang District Beijing China 100022 Page 12 Figure5 7 2 Select the serial of configuration of serial software Port number you can select port Number from device manager of your system Set the baud rate 115200 DF702 is 115200 baud rate and DO100 is 9600 Parity bit NONE Data bits 8 Stop bit 1 As shown in figure 5 8 you can follow the given instruction ...

Страница 13: ... to the serial software such as devaddr nwkskey etc please refer to the specific instruction format LoRa instruction manual Sending First Command send input command AT ID read all as shown in figure5 8 Response of Device ID DevAddr 01 7D 53 74 ID DevEui 47 68 B2 69 00 2A 00 43 ID AppEui 52 69 73 69 6E 67 48 46 Sending Second Command AT KEY NWKSKEY 6132f00ad1efd2d4af5e12f8ef745d8b set nwkskey as sh...

Страница 14: ...eg AT ID DevAddr 01234567 Set DEVEUI AT ID DevEui 8 bytes length hex identifier 64bits eg AT ID DevEui 0123456789ABCDEF Set NWKSKEY AT KEY NWKSKEY 16 bytes length key eg AT KEY NWKSKEY 2B7E151628AED2A6ABF7158809CF4F3C Set APPSKEY AT KEY APPSKEY 16 bytes length key eg AT KEY APPSKEY 2B7E151628AED2A6ABF7158809CF4F3C Detailed instructions please refer to the RHF PS01509 LoRawan Class AC AT Command Sp...

Страница 15: ...ection is wrong there will be no reply sending instructions Confirm the baud rate is the correct choice of baud rate the correct should be 115200 5 4 Video link Configuration of Lora parameter DF702 Smart Trash Bin Sensor is shown with detailed you can follow these instructions as well as click at video link https youtu be nklTd88SPXI 6 Configurations of Relevant Parameters of Device Through the c...

Страница 16: ...s Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 16 Figure6 2 Serial software 6 2 Connect TTL to sensor Device serial port GND TX and RX positions have been marked on the diagram Upper diagram the connection of RX TX and GND is given with detailed if you not connect properly then you cannot configure your device properly ...

Страница 17: ...oad 12 Chaoyang District Beijing China 100022 Page 17 Figure6 3 TTL and Device ports connection are shown in the figure 6 4 Figure6 4 Steps 1 Confirm that the jumper cap is at the solid position jumper cap defaults to solid position as shown in figure6 5 Figure6 5 2 Connect TTL to sensor as shown in figure6 4 and figure6 6 Jumper cap is in solid position ...

Страница 18: ...ure6 6 6 3 Use serial software to configure Open the serial software Select the serial port number Set the baud rate 115200 Parity bit NONE Data bits 8 Stop bit 1 Connect the device s power When you see the serial port output sending instructions are in Hexadecimal then we will change the format of serial software sending option Data is sent in hexadecimal format please refer to instruction format...

Страница 19: ...rvice dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 19 Figure6 7 Note Before sending the command please first click the enter and then click Send Example 1 change upload time to an hour the instruction is 99990101 as shown in Figure 6 8 ...

Страница 20: ... of the corresponding equipment Notes 1 When the device is at Sleep mode sending instructions will be invalid 2 Before sending instructions please click enter and then send 3 The command is sent in hexadecimal format except for configuring APN for details refer to the protocol 4 Check whether the RX and TX of the TTL and the device are correctly connected If the connection is reversed no output is...

Страница 21: ...g 7 Upgrade devices 7 1 Instructions Dear customer if you want to upgrade the files you need to contact us to gets information about new version and use our software to upgrade the files If device has special requirements or needs to be customized please contact sales department for confirmation If there is no special features it is not recommended to upgrade the equipment 7 2 Tools A TTL and cabl...

Страница 22: ...ounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 22 TTL and device connection as shown in figure 7 3 Figure7 3 Confirm that the jumper cap is in the solid position jumper cap defaults to solid position as shown in figure7 4 Figure7 4 3 Connect TTL to sensor as shown in figure7 3 and figure7 5 ...

Страница 23: ...Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 23 Figure7 5 7 4 Upgrade steps 7 4 1 Steps Up gradation of device can be possible to following the steps Open the serial software select the desired serial port set the baud rate 115200 parity None data bits 8 stop bit 1 ...

Страница 24: ... dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 24 Figure7 7 7 4 2 Step Click the Open button enter in the input box click the Enter as shown in the figure 7 7 set the time interval to 500mm or 500mm check the cycle to send options click send ...

Страница 25: ...Dingtek Technology Corp Ltd Tel 86 10 64718863 64753700 http www dingtek com E mail service dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 25 Figure7 5 ...

Страница 26: ...ek com E mail service dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 26 Figure7 6 7 4 3 Restart the device you can see as given Figure 7 6 Next input is u click on the line and Click send you can gets output result ccccccc ...

Страница 27: ...ngtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 27 Figure7 7 7 4 4 When you receive Output result ccccccc then click at close of this Serial port software 7 4 5 Before open the Hyperterm exe file please close the Serial software it is necessary step ...

Страница 28: ...software Hyperterm exe as shown in Figure7 10 dialog box enter 1 in the name bar and click OK Figure7 11 You can see Figure 7 12 dialog box select the desired serial port in Figure selected port is such as COM16 and then click OK The Figure7 13 dialog box appears selects the following steps in dialog box Select baud rate 115200 Parity None Data Bits 8 Stop Bits 1 Data Flow Control bits 1 After sel...

Страница 29: ...k Technology Corp Ltd Tel 86 10 64718863 64753700 http www dingtek com E mail service dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 29 Figure7 10 Figure7 11 ...

Страница 30: ...63 64753700 http www dingtek com E mail service dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 30 Figure7 12 7 4 7 Click the connection option you can receive output cccc on the screen Figure7 13 ...

Страница 31: ...ice dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 31 Click at send option and select the file which you want to upgrade select the address of your file Agreement selecting option you can select given modem Ymodem and Click Send Figure7 14 ...

Страница 32: ...nfirm the results of figure7 6 on the serial port software please checks connection of RX TX and TTL device 7 5 Video link The instruction of upgrading the device is given with detailed this link can help you to upgrade the device corresponding the different correct steps DF702 Smart Trash Bin sensor Upgrade Instruction https youtu be zUQbWT6V414 Attachment 1 Network server For the network Server ...

Страница 33: ...er During the usage of CNDingtek Network Server you need no more settings just connect the device with power and use it easily For network server you can contact to customer Services E mail service dingtek com 1 2 TTN Network Server After login Register Gateway Click Gateway register gateway Select I m using the legacy packet forwarder fill in the registration information click Register gateway to...

Страница 34: ... 64753700 http www dingtek com E mail service dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 34 Click Devices register device to input device information DevAddr Deveui and APPKEY and click Register ...

Страница 35: ...http www dingtek com E mail service dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 35 1 3 Loriot login login interface Log in China please choose CN1 Shenzhen Chin Enter the user name and password click login ...

Страница 36: ...ss Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 36 Add Applications 1 3 2 Add application Login you can see Gateways and Application information Click Application to add an application 1 3 3 Add device Click Dashboard Applications DF702 Added Applications Manage Device to add devices There are two ways to add a device ...

Страница 37: ...under Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 37 The Loriot server automatically generates OTAA mode devices so parameters such as Deveui Appeui Appkey are generated by the server If the device is automatically generated by loriot you need to modify the corresponding parameters of the device ...

Страница 38: ...a 100022 Page 38 Import devices Import ABP type devices Need to add DevEUI DevAddr NwkSKey AppSKey FcntUp FcntDn Label already contains DevEUI DevAddr NwkSKey AppSKey information FcntUp FcntDn may be 0 number is possible Import OTAA type devices Fill in DevEui AppEui and APPKey and then click Import device To add success Device label are already contains DevEui AppEui and APPKey information ...

Страница 39: ...ervice dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 39 1 3 4 Add Gateway Select the type of gateway used Follow the prompts to add gateway information Note loriot generates a gateway ID based on the MAC information of the added gateway ...

Страница 40: ...64718863 64753700 http www dingtek com E mail service dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 40 Take as an example Need to enter the gateway MAC and location and then click register ...

Страница 41: ...ge 41 Click on the registered gateway you can see the gateway information and can choose the required band such as US915_CH0 In the gateway information interface you can also monitor the frequency of the selected band as shown below According to this frequency check the frequency of the server gateway and device frequency point three frequency is the same If not the data can not be uploaded correc...

Страница 42: ...e dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 42 1 4 Actility 1 4 1 Add ABP mode device Enter the relevant information as shown below Device name Custom Port Number 3 1 4 2 Add OTAA Mode Device Enter the relevant information as shown below ...

Страница 43: ...ounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 43 1 4 2 Add a gateway Note Actility supports lrr mode gateways and does not support packet forwarding mode Given picture you need to add these contents make sure of New Base Station 2 Common problem Facing by the Customers and theirs solutions ...

Страница 44: ...ode the deveui appeui should be same as the registration Shorten distance between sensor and gateway Some LoRa server like TTN the things network ask for special devaddr Please use TTL UART to connect with sensor and change the parameter Please check whether the gateway is correctly connected to the network server Please check the position of the jumper cap whether the jumper is on the solid posit...

Страница 45: ...pletely blocked Device have Fall Function The standard version will measure with full fire and battery check features and enhanced version will measure full fire and battery check and fall function Can I change my device data interval to 5 minutes or 1 minute Can be changed to 5 minutes can not be changed to 1 minute the device default 10 minutes test once you can change it to 5 minutes the operat...

Страница 46: ... http www dingtek com E mail service dingtek com Room A209 Flounder Business Park Shunbai Road 12 Chaoyang District Beijing China 100022 Page 46 Time State Upload data or no 10 10 full Yes first data 10 20 full No 10 30 not full Yes 10 40 not full No 10 50 full Yes ...