GETTING FAMILIAR WITH THE GTS-1000 CAPABILITIES 

 
Working area 

The GTS-1000 Series features a large working area ideal for testing large samples up 

to 

10.2 inches (260 mm) wid

e.  The versatile T-slots can secure unusual shaped fixtures 

or parts in tension.  T-nuts and screws may be used to secure other fixturing. 
 
Do not allow fluid to run into the T-slots.  Tests with potential for running fluids should 
secure The Dillon oversized plate accessory (available separately) to the T-slot work 
surface and remove the side T-slot cover plates.  Always unplug the stand immediately 
if fluid runs into the T-slot and do not reconnect until all fluid is cleared and dried. 
 

WARNING 

 

Do not allow fluid to run into the T-slots.  This can 

produce an electrocution hazard.  Always unplug the 
stand immediately if fluid runs into the T-slot and do 

not reconnect until all fluid is dried.

 

 
 
 
 
The GTS-1000 includes a handy tension base plate for in-line tension 
grips as well.  To remove or install, remove one cover plate side 
screw and loosen the other.  Rotate cover 90 degrees. Slide tension 
plate with T-nuts loosely attached in or out as desired.  Rotate cover 
back and retighten screws.  
 

Column 

The GTS-1000 Series is a single column stand designed for vertical use.  Maximum 
crosshead trav

el is 18.7 inches (475 mm).  

 
The GTS offers exceptional working headroom of 22.8 inches (580 mm) with the GTX or 
GS force gauge.  Over 5 inches (125 mm) of extra headroom reduces travel loss impact 
due to fixturing.  Extension rods may be required for short samples with low profile 
fixturing.  In compression, blocks may be used to elevate the sample as needed.   
 

Содержание GTS-1000

Страница 1: ...GTS 1000 Digital Control Motorized Test Stand Operator s Manual ...

Страница 2: ...to power 7 Selecting units of measure for speed 7 Alignment 7 Mounting the force gauge 8 Getting Familiar with the GTS 1000 Testing Capabilities 9 Working area 9 Column 9 Crosshead 10 Connectors 10 Operation 10 Front console 10 Emergency stop button 12 Travel limits 12 Power switch 13 Accessories 13 Performing a Test 14 Troubleshooting 15 Specifications 16 2 3 07 Rev B ...

Страница 3: ...ed by a Dillon force gauge grips and accessories it constitutes a key component in force measurement systems to accurately test a wide range of products This tester may be used for tensile and compression tests of many types e g break peel flexural top load and many more The digital controls make testing speed more accurate and the stand more intuitive to operate The crosshead features he Dillon G...

Страница 4: ...onnel RISK OF ELECTRICAL SHOCK Do not allow fluid to run into the T slots This can reach the electrical circuitry Tests with potential for running fluids should have a larger plate secured to the T slot work surface Always unplug the stand immediately if fluid runs into the T slot and do not reconnect until all fluid is cleaned Verify Input Power Source BEFORE operation The GTS 1000 may operate wi...

Страница 5: ...nment Be sure to align any lower fixture location in line with the attached force gauge stem The force must be applied in line for tests to be accurate and to avoid gauge damage Understand all aspects of testing equipment operation before operating See users manual or contact your Dillon dealer with any questions 5 ...

Страница 6: ...key for tension plate hardware Two screws to mount gauge If any damage is discovered do not go any further with installation and do not connect the unit to a power supply under any circumstances We recommend you keep all packaging for future shipping requirements Setting up for your supply voltage The GTS 1000 Power Module has two settings 110 120 VAC 220 240 VAC The GTS 1000 requires a stable reg...

Страница 7: ...cks Selecting units of measure for speed The GTS 1000 displays speed in inches per minute or millimeters per minute as shown by the LED annunciator To toggle the units press and speed keys simultaneously Although easily changed this parameter is generally set only once by most users Alignment Proper alignment is important for testing accuracy and to prevent damage to the stand gauge and fixtures 1...

Страница 8: ...er force gauge will be mounted discuss mounting with your Dillon dealer A with the tester columns Use included screws to secure If CAUTION When using a force than a Dillon GS or gauge other GTX it is the installer s responsibility to ensure 1 strength of all connections and insure hardware mounting depth falls within gauge manufacturer requirements and 2 load cell stem columns within is centered o...

Страница 9: ...slots This can produce an electrocution hazard Always unplug the stand immediately if fluid runs into the T slot and do not reconnect until all fluid is dried The GTS 1000 includes a handy tension base plate for in line tension grips as well To remove or install remove one cover plate side screw and loosen the other Rotate cover 90 degrees Slide tension plate with T nuts loosely attached in or out...

Страница 10: ...rs are on the rear of the stand Position Use Top n c No connection Factory and dealer use only Middle GTX Connect to a Dillon GTX force gauge for stand reverse stop control on sample break or compression overload Bottom PC Sends the GTX serial information to a computer when the stand reverse stop cable above is used The PC port eliminates Y cables or expansion boxes required by other brands OPERAT...

Страница 11: ...ts The Dillon GTX force gauge offers increased stand control In Manual mode the GTX can stop the stand upon sample break detection In Auto reverse it can reverse the stand upon sample break detection The GTX also interrupts downward travel when gauge capacity is exceeded to reduce potential for overload See accessories for the GTX stand reverse cable CAUTION The overload detection within the GTX i...

Страница 12: ...rgency stop button Emergency stop button The GTS 1000 Series features an emergency stop button conveniently located at the top of the console This can be used as a panic switch whenever conditions are becoming a concern Engaging the emergency stop is also a good precaution to avoid inadvertent crosshead movement when the stand is unattended When emergency stop button is depressed motor control is ...

Страница 13: ...e power entry module at the rear of the stand When turning the machine OFF it may take about 15 seconds for the speed indicator display to extinguish while the internal capacitor self discharges Do not turn power back on until the front panel lights have extinguished NOTICE The console lights require about 15 seconds to completely discharge when the power is turned OFF ACCESSORIES Force gauges man...

Страница 14: ...ent test starting point k Note the starting crosshead measurement on the ruler l Adjust test speed m Check manual or auto reverse mode n Start test with or Ø key o Note crosshead measurement at point of interest p Save print Q Graph report or send peak gauge reading to Q Data q Crosshead Return for new sample start again at step f Some steps are optional depending upon test type and the informatio...

Страница 15: ...e units of speed Stand does not reverse at limit Manual mode selected Change mode to auto reverse No reverse upon sample break with GTX Prior test peaks not cleared Press RESET on the GTX prior to starting each test GTX configuration Set break detect and verify peak reading is larger Gauge not in alignment with fixture Alignment adjustments See Alignment section of manual Grinding noise Non smooth...

Страница 16: ...consumption 150 watts maximum Stand weight 54 lb 25kg Travel range 18 7in 475 mm Maximum headroom 22 8 in 580 mm Working area 10 2 x 10 2 in 255 x 255 mm Tension plate Three 10 32 UNF holes Speed range 0 2 40 0 in min 5 1000 mm min Speed display Digital LED Direction of travel indicated on stand Yes Operating modes Manual and auto reverse Reverse on sample break Yes with GTX and cable Reverse on c...

Отзывы: