DIESTRO 103APV4R ELECTROLYTE ANALYZER
83 JS Medicina Electrónica
3. REVIEW AND EVALUATION OF THE STATUS OF THE ELECTRODES
When there is a solution in the measuring chamber, the analyzer can measure the voltage (mV)
in the electrode train, proportional to the concentration of each electrolyte.
The observed voltage allows the operator to evaluate:
•
Electrode Stability: Voltage without significant variations implies stable electrodes (it is
necessary to wait a stabilization time -15 seconds- from when the sample is loaded).
•
Electrode Gain: Differences between the voltages generated by the StdA and the StdB.
In some cases it may be useful to determine the voltage generated by certain samples in the
measuring chamber.
The positioning of each calibrator solution or sample in the measuring chamber can be done
both manually (manual window) and automatically (automatic window). Please note that the
standards that are consumed from the pack in Manual Mode will not be discounted, so the Pack
will actually have less liquid than the one reported.
Содержание 103APV4R
Страница 1: ...ELECTROLYTE ANALYZER OPERATOR AND SERVICE MANUAL DIESTRO 103APV4R V 1 0 MARCH 2022 RE 0851 CONTENT ...
Страница 91: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 91 JS Medicina Electrónica Needle sampling with Tip placed ...
Страница 94: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 94 JS Medicina Electrónica 6 Put the needle cover and screw it ...
Страница 109: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 109 JS Medicina Electrónica Pack Rear View Cable Uchip ...
Страница 111: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 111 JS Medicina Electrónica Autosampler Pack ...
Страница 112: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 112 JS Medicina Electrónica Rear view detail ...
Страница 113: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 113 JS Medicina Electrónica Front View with Open Lid ...
Страница 114: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 114 JS Medicina Electrónica Electrode cable Electrode lock Electrode chamber ...