2. ACCESS TO THE SERVICE MENU
2.1.
To access the manual maintenance menu, navigate to:
Route
: "HOME" screen → → Maintenance → Manual
To manually position StdA, press the "A" button and then CW (with General Persitáltica
enabled), until the liquid is brought to the desired position.
By pressing "Read" you will be able to see the conductivity in the selected detector (see map)
A: StdA
B: StdB
C: Sample by FP/Sampler
It is not advisable to open more than one valve at the same time as this can
cause air to enter the hydraulic circuit. If necessary, purge the analyzer after
testing.
Do not keep the valves open for too long.
Pressing Esc. does not close the open valves.
To modify threshold values, navigate to:
Route
: "HOME" screen → → Settings Menu → Detection Thresholds
To be handled only by trained technicians. This will affect the circulation of
liquids in case it is improperly modified.
Содержание 103APV4R
Страница 1: ...ELECTROLYTE ANALYZER OPERATOR AND SERVICE MANUAL DIESTRO 103APV4R V 1 0 MARCH 2022 RE 0851 CONTENT ...
Страница 91: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 91 JS Medicina Electrónica Needle sampling with Tip placed ...
Страница 94: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 94 JS Medicina Electrónica 6 Put the needle cover and screw it ...
Страница 109: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 109 JS Medicina Electrónica Pack Rear View Cable Uchip ...
Страница 111: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 111 JS Medicina Electrónica Autosampler Pack ...
Страница 112: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 112 JS Medicina Electrónica Rear view detail ...
Страница 113: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 113 JS Medicina Electrónica Front View with Open Lid ...
Страница 114: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 114 JS Medicina Electrónica Electrode cable Electrode lock Electrode chamber ...