![DiaSys O-Maxi OM Series Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/diasys/o-maxi-om-series/o-maxi-om-series_operators-manual_2494672013.webp)
O-Maxi operator’s manual Version: 2020.02.10
12
4.5
I
NSTALLATION OF THE WATER PURIFIER MEMBRANE
1.
Disconnect the water inlet pipe from the membrane holder (Figure 1)
2.
Remove the membrane holder out of its two plastic clips (Figure 2)
3.
Unscrew (by turning to the right) the cap of the membrane holder using the two special keys
provided with the accessories.
4.
Once the membrane cap is removed, insert the new membrane (Figure 3), peripheral seal at
the top, to the complete stop: the end of the collecting tube must be flush with that of the
membrane holder. (Figure 4)
figure 1
figure 2
figure 3
figure 4
5.
Check that the O-ring is well-positioned at the bottom of the retaining wall of the membrane
holder (see the plan).
6.
Unscrew (by turning to the right) the high streaked part ("big cap") of the membrane holder.
7.
Joint the membrane on its stirrups.