Op
erator’s manual
version: 2021.06
30
Understanding conductivity values
6.4.2.1
General information
The global quantity of dissolved solids in water can be measured by the CONDUCTIVITY (expressed in
microSiemens per centimeter (µS / cm).
6.4.2.2
Production conductivity « output of water purifier »
Class 2 water (see ISO 3696 standard)
Convenient for clinical analyzers.
The conductivity of the
produced water is below
1µS/cm
The conductivity indicator
is switched OFF
Class 3 water (see ISO 3696 standard)
If the desired water class is 2:
1.
Replace the resin (see paragraph
2.
Follow the instructions of paragraph
HI
(too high) is displayed
when the conductivity is
between 1.1 µS/cm to
5µS/cm
The conductivity indicator
is ON with yellow colour
1.
Replace the resin (see paragraph
2.
Follow the instructions of paragraph
SAT
(saturated) is
displayed when the
conductivity is above
5µS/cm
The conductivity indicator
is ON with red colour
When the conductivity is
above 9.9,
++
is displayed
0.1
US/CM
READY
1.5
US/CM
HI
READY
5.5
US/CM
SAT
READY