![Diamond SmartLyte Скачать руководство пользователя страница 61](http://html.mh-extra.com/html/diamond/smartlyte/smartlyte_service-manual_2493455061.webp)
6 Adjustments
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 61 of 81
6.
Adjustments
The SMARTLYTE Electrolyte Analyzer has been designed to require minimal electronic
adjustment. Diamond Diagnostics recommends adjusting the sample sensor circuit after
changing the sample sensor during routine troubleshooting or during annual maintenance
procedures. The temperature adjustment is active and requires the temperature adjustment
check plug for correct calibration of the temperature circuit.
6.1.
Sample Sensor Adjustment
Press F7 in the keyboard to access
SERVICE MENU
and here you need to access
SAMPLE SENSOR CHECK
. By confirming, instruments will remove solutions from
sampling path. On the top right corner of the screen, a number between the ranges of 0 to
240 is shown. Sample sensor potentiometer (R189) is located on the top of the CPU
board (See Fig. 6-1).
Adjust R189 until the value 100 is reached in the display. After this adjustment, open
sample door, press
YES
or
ENTER
, and introduce water in the sample probe. When
liquid arrives in the sample sensor, the number on the screen will change. Displayed
values must be 50 points or higher that the adjusted value (80-120 range).
Note: R189 can be accessed through the rear cover vent slots.
Fig. 6-1
R189 S.S.
ADJ.
Содержание SmartLyte
Страница 1: ......
Страница 11: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 11 of 81 1 OVERALL FLOW ...
Страница 12: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 12 of 81 2 MAIN MENU HUMAN ...
Страница 13: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 13 of 81 3 MEASUREMENT SEQUENCE ...
Страница 17: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 17 of 81 6 MAINTENANCE ...
Страница 22: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 22 of 81 9 SERVICE MENU ...
Страница 25: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 25 of 81 11 POWER UP ...
Страница 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Страница 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Страница 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Страница 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Страница 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Страница 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Страница 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...
Страница 81: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 81 of 81 9 3 Display 9 4 Valve ...