![Diamond SmartLyte Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/diamond/smartlyte/smartlyte_service-manual_2493455027.webp)
3 Software Operation
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 27 of 81
3.2.
Service Menu
The SERVICE MENU (mode provides a menu of service check functions to assist in
failure diagnosis and identification. A logic flow chart is provided for reference of the
sequences included in the SERVICE MENU mode.
CHECK ELECTRODES
This mode allows for the measurement and display of electrode voltages.
STD A, STD B
or an external sample can be measured to determine electrode performance and provide
measurement data useful in locating electrode calibration faults. Each electrode
configuration will be displayed and the electrode voltage associated with the sample type
selected can be observed.
Example Displays:
Std. A: Na=1582 Std. B: Na=237
K= 654 Cl=-904 K=-988 Cl=647
Ext: Na=1321
K=-598 Cl=-127
CHECK SAMPLE SENSOR
In this mode, the sample sensor output is displayed which is useful for electrical
adjustment or for troubleshooting sample sensor faults. A dry sample sensor should
display a value of approximately 100 with a range of 80 to 120. After verifying a dry
sensor, aspiration of a clear fluid (i.e., water) should increase the displayed value by at
least 40 units.
Example Display:
DRY 80 -120: 102 External: 153
YES=Pump/No=Exit Pump: NO=OFF
Содержание SmartLyte
Страница 1: ......
Страница 11: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 11 of 81 1 OVERALL FLOW ...
Страница 12: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 12 of 81 2 MAIN MENU HUMAN ...
Страница 13: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 13 of 81 3 MEASUREMENT SEQUENCE ...
Страница 17: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 17 of 81 6 MAINTENANCE ...
Страница 22: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 22 of 81 9 SERVICE MENU ...
Страница 25: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 25 of 81 11 POWER UP ...
Страница 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Страница 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Страница 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Страница 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Страница 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Страница 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Страница 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...
Страница 81: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 81 of 81 9 3 Display 9 4 Valve ...