1 Introduction
ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 32 of 58
Figure 1-45 Valve positions
Solenoid Valve assembly removal requires the removal of the indicator cap and crimp pin first.
Proceed as follows;
1. If the instrument is on, place in Hold in Maintenance mode to prevent reagents from cycling
and valves from activating.
(
Main Operations Menu
>>>
Maintenance
>>>
Hold in Maintenance
)
2. Locate the opening of the indicator cap where the solenoid shaft is visible. Orient opening
upwards.
Figure 1-46 Valve positions
3. Push down on Indicator cap in the direction of the arrow to slide it off the solenoid valve shaft.
4. Then, remove the Crimp Pin from the solenoid shaft.
To install the crimp pin and cap, the instrument must be on.
1. Place the instrument in Hold in Maintenance mode.
(Main Operations Menu>>>Maintenance>>>Hold in Maintenance)
2. Access the Check Valve menu in Diagnostic Menu.
(Main Operations Menu>>>Diagnostics>>>Check Valves)
3. Press the button for the valve that needs the Crimp Pin and Indicator Cap installed.
(Check Valves>>>Check Valve A (B, C, R or V)
4. This energizes the solenoid and the shaft is extended to the outer most position
Indicator Cap
Opening
V Valve Solenoid
Screws