Document Title:
Remote Tomo5 operation guide for eBIC microscopes
Author(s):
Kyle Morris
Julika Radecke
Andrew Howe
Version:
2.0.3
Date:
27 Jun 2022
If you have any questions at any point during set-up, please ask your Local Contact for assistance Page 24 of 29
5 Final checks and data collection
5.1 Auto Functions and alignment
5.1.1 Alignments via Autofunctions
Navigate in the Atlas (refer to 3.1.3) to an area of carbon to perform alignments
You must be at eucentric height (refer to 3.2.2) and the objective aperture should be removed
Use the Auto Functions in the order as below:
-
The order should be followed, if Auto Function is successful then move on to the next
-
Ensure that you select the corresponding Preset for each Auto Function alignment
1. Auto Functions
→
Auto-Functions (TEM)
→
Auto-eucentric by stage tilt
→
Start
Presets -> Eucentric Height for Auto-eucentric by stage tilt or beam tilt
2. Auto Functions
→
Auto-Functions (TEM)
→
Autofocus
→
Start
Presets
→
Focus for Autofocus
3. Auto Functions
→
Auto-Functions (TEM)
→
Autostigmate
→
Start
Presets
→
Thon Ring for Autostigmate
4. Auto Functions
→
Auto-Functions (TEM)
→
Autocoma
→
Start
Presets
→
Thon Ring for Autocoma