background image

 

 

 

Installation of Regulator Kit 

onto 1/4 in. molstic-S™ 

Mounting Platform 

P/N 401880  

Instruction Sheet

 

This kit is used to upgrade 1/4 in. single or dual 
molstic-S platforms that do not include the integrated 
pressure regulator.  This upgraded configuration is 
ideal for the situation where the device under test 
(DUT) will be tested while plumbed downstream of the 
molbloc-S. 

 

  High pressure gases are potentially hazardous.  

Energy stored in these gases can be released 
unexpectedly and with extreme force.  High 
pressure systems should be assembled and 
operated only by personnel who have been 
instructed in proper safety practices.

 

 

 

Figure 1.

  1/4 in. molstic-S, with Regulation 

PREPARING THE REGULATOR FOR 

INSTALLATION 

 

Place a new VCR

®

 gasket (P/N 102183) into the inlet 

port of the regulator.

 

 

Place the VCR male nut (3) over the tube adaptor (2) 
and thread it into the regulator inlet port and tighten 
finger-tight. 

 

  Over-tightening VCR connections will damage 

the sealing beads and possibly cause system 
leakage.  Do not re-use metallic VCR gaskets. 

 

 

Hold the regulator stationary, and tighten the male nut 
1/8 turn past finger-tight.

 

 

Install the quick connect body (1) onto the tube fitting 
and tighten 1-1/4 rotations beyond finger-tight.

 

 

Place a new VCR gasket into the outlet port of the 
regulator.

 

 

Thread the VCR male union (4) into the regulator outlet 
port and tighten finger-tight.

 

 

Hold the regulator stationary, and tighten the male union 
1/8 turn past finger-tight.

 

 

Attach the regulator mounting bracket to the base of the 
regulator using the two included #10-32 socket head cap 
screws, #10 flat washers, and the #10 lock washers.  The 
long side of the bracket should be perpendicular to the 
direction of flow through the regulator. 

INSTALLATION OF REGULATOR ONTO 

MOLSTIC-S 

 

Using a pair of properly sized wrenches remove the 
VCR x tube adaptor which is on the inlet end of the 
molstic-S.  This will leave a VCR coupler (5) as the 
molstic interface.  Retain the adaptor for future use.  
The VCR gaskets are for one-time use and should 
therefore be discarded.

 

 

Using a pair of properly sized wrenches remove the VCR 
coupler  and filter as an assembly (5 and 6) from the 
female VCR swivel nut (7).  Discard the VCR gasket.

 

 

Place a new VCR gasket securely against the sealing 
surface inside the female VCR coupler (5) of the 
molstic-S inlet fitting.

 

 

Tighten the VCR coupler and filter subassembly to the 
regulator outlet fitting 1/8 turn past finger-tight.

 

 

Place a new VCR gasket securely against the sealing 
surface inside the female VCR swivel nut (7).

 

 

Place the regulator with its bracket on top of the molstic-S 
platform.  Align the filter assembly’s male VCR outlet 
fitting with the mating female VCR swivel nut and thread 
the nut onto the fitting.

 

 

Hold the filter 

assembly

 stationary with a backup 

wrench.  Tighten the female nut 1/8 turn past finger-
tight.

 

 

Figure 2.  

molstic-S Nut Plate 

 

 

Отзывы: