Calibration Solutions
for Pressure
and Flow™
PK-7000-MPC1-D
Interconnections Kit
Installation Instructions
INTERCONNECTIONS KIT, P/N 401645, INCLUDES
•
6 100319 Adaptor, 1/8 in. tube x 1/8 in. NPT M
•
2 102184
Reducer, 4P x 2P
•
2 100317 Adaptor, 1/8 in. tube x 1/4 in. NPT F
•
1 101988-Z Tubing, SS, .125 in. OD x 300 cm
•
2 401459 Quick Connect and Stand Assy
•
2 102116
Quick Connect, 1/8 in. NPT F
PARTS ON TEST PORT OR IN ACCESSORIES OF PG7000
•
1 101200 Collar
•
1 400518 Adaptor , 1/8 in. tube x DH 200 M
•
1 100272 Gland
TO PNEUMATICALLY CONNECT THE PG7000
S
OR RPM
S
TO THE MPC1-D
•
MPC1-D is intended for controlling pressures in a high line differential pressure system where there is a “high” side and
a “low”. The instructions below should be repeated, once for the high side and once for the low side. See the MPC1-D
Operation and Maintenance manual for a schematic of a typical system layout.
•
If you are using RPMs (not PG7000s) go directly to Step
p
b.
n
Remove gland, collar and plastic plug from the PG7000 TEST port.
o
Install gland and collar onto 1/8 in. tube x DH200 male adaptor supplied in PG7000 accessories.
Note that the collar threads left handed. Assure it is properly attached. The orange plastic plug is
not used.
p
a
PG7000 - Screw the adaptor (as assembled) into the PG7000 TEST port. Torque to 15 Nm (Newton
meter).
p
b
RPM3 - Screw 1/8 in. tube x 1/8 in. NPT male adaptor into the RPM3 TEST port using Teflon tape as
a seal.
TO MPC1-D
TEST HI or TEST LO
TO MPC1-D
REF HI or REF LO
101988-Z
102184
100319
100319
100317
102116
401459
101988-Z
400518
INCLUDED WITH P 7000
G
100272
100319
TO RPM3
101200
TO P 7000
G