GFS2102™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 82
FEATURE
DESCRIPTION
Save AutoZero
information to file
(check box)
All information used to determine the AutoZero correction is saved to a delimited text file when
this option is selected. The resultant file is used to validate and quantify the corrections
determined by the AutoZero process. All AutoZero files are automatically saved in the
\AutoZ
directory relative to the root data file directory selected in
[Tools][Options][Data File]
and
have an “.atz” extenstion.
The file is not handled by
GFS Tools
. Analysis of the AutoZero files requires a separate
spreadsheet application such as MS Excel.
Save SDF file for
each accumulation
(check box)
The SDF file (Sample Data File) provides a complete account for all outputs and calculations
used to determine flow from the GFS and DUTs. Check this option to create a SDF file for
each flow accumulation. Use this option to troubleshoot problems or to verify
GFS Tools
calculations.
When this option is selected, the SDF file for the GFS is automatically saved in a
\SDF
directory relative to the root data file directory selected in
[Tools][Options][Data File]
. SDF
files for a DUT are saved in the same directory as the DUT data file and have an “.sdf” file
extension.
SDF files are not manipulated by
GFS Tools
. Analysis of the SDF files requires a separate
spreadsheet application such as MS Excel.
Allow manual GFS
manipulation when
running a Test
Definition
(check box)
Check this option to allow a user to use the run time toolbars on the GFS Run Screen and
other device run screens when running a Test Definition. By default this option is not
selected. The choices selected in the Test Definition govern the test process. However, there
are times in which manual user interaction is more efficient. For example a user may
manually adjust the target flow or AutoZero the GFS when a test is active.
Delay timed point
collection until GFS
takes a data sample
(check box)
Check this option to allow
GFS Tools
to trigger timed point sampling only after the GFS has
started to sample data. This choice affects manual tests only. By default this option is
selected.
Depending on the sample setup, a significant amount of time may be required to collect the
first sample. Un-check this option to log data during this pre-sample period. Although a GFS
flow will not be available, valuable information on the changes in ambient conditions and the
AutoZero correction can be used.
Abort timed point
collection when
accumulation is
aborted
(check box)
Check this option to allow
GFS Tools
to automatically stop timed point collection when the
GFS accumulation is complete. This option affects manual tests only. By default this option
is selected, however it may be desirable to uncheck this option to log additional ambient
conditions information after the accumulation is complete.
Abort flow control
when accumulation
is aborted
(check box)
Check this option to set the flow to zero after the accumulation is complete. This option
affects manual tests only. Do not select this option when multiple accumulations will be used
at the same flow rate. The intended use of this option to save cylinder gas when a long
duration unattended accumulation is performed.
Abort accumulation
after target mass is
reached
(check box)
Check this option to end the flow accumulation after the target mass is reached. If the option
is not checked, user intervention is required to stop the accumulation. This option affects
manual tests only. By default this option is checked.
Содержание GFS2102
Страница 1: ...2007 DH Instruments a Fluke Company GFS2102 Gravimetric Flow Standard Operation and Maintenance Manual...
Страница 8: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 14: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XII N NO OT TE ES S...
Страница 16: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XIV N NO OT TE ES S...
Страница 20: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 4 N NO OT TE ES S...
Страница 32: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...
Страница 58: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 42 N NO OT TE ES S...
Страница 174: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 158 N NO OT TE ES S...
Страница 182: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 166 N NO OT TE ES S...
Страница 204: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 188 9 2 TYPICAL TEST SETUPS...
Страница 206: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 190 N NO OT TE ES S...