GFS2102™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 76
Table 23. <molbox Output>
Run Screen Fields
LABEL
DESCRIPTION
Red/Green Circle
molbox flow Ready/Not Ready indicator (see the molbox Operation and Maintenance
Manual, Ready/Not Ready Section). The circle is green when molbox flow is ready
and red when not ready.
Rate
Instantaneous molbox flow rate of change in unit of flow per second.
Flow
molbloc flow in the current test unit of measure.
Reynolds Number
Reynolds number of the flow through the molbloc as calculated by the molbox.
Upstream Pressure
Pressure read by the molbox upstream pressure transducer.
Downstream
Pressure
Pressure read by the downstream molbox pressure transducer.
Differential Pressure
The pressure drop across the molbloc that is currently being used by the molbox to
calculate flow.
The differential pressure is the difference between the molbox upstream and
downstream RPTs (in molbox RFM, it may be the measurement made by the
microrange RPT). The molbox determines the differential pressure and outputs it to
GFS Tools
..
BPR
BPR is the back pressure ratio, downstream pressure/upstream pressure. This field
is present only when the DUT is a molbloc-S. The value displayed is either the
measured ratio or N/A when the value is not measured. The molbox front panel BPR
setting determines whether or not BPR is measured or if the 2 molbox RPTs are
averaged. Refer to the molbox operation and maintenance manual for more
information.
Temperature
The molbloc temperature as calculated by the molbox from the average of the molbloc
upstream and downstream PRT’s.
Set
Active molbox flow set point. This value is used only when the molbox is equipped
with the MFC control option and is used as the analog supply to a flow controller. For
manual tests change target flow set point by using the
GFS Run Screen
. For trouble
shooting purposes, change the set point by entering a new set point value and
pressing
[Enter]
.
Output
Active molbox analog output. As with the
<Set>
, this value is used only when the
molbox is equipped and used in to support an analog flow controller or DUT.
Switchbox
molbox MFC switchbox control and measurement channels. The format is
<Set
Channel . Measurement Channel>
. This field has no value if the molbox does not
have the MFC control option and an MFC switch box is not in use.
K Factor
Active DUT gas conversion value in use by the molbox. The value is specified during
Test Initialization. Change the value in Manual Test operation by entering a new K
Factor and pressing
[Enter]
. See the molbox Operation and Maintenance Manual, K
Factor Section, for details on the use of K factors.
Gas
Active molbox calibration gas. When running a Manual Test, this value can be
changed using the toolbar at the bottom of the display. In all other run modes, the test gas
is selected during test initialization.
Stability Setting
Active Ready/Not Ready stability criterion in the molbox. This value is fixed by the
Test Definition in use when in Run Test mode. In all other run modes, the stability
setting can be changed by entering a new stability setting and pressing
[Enter]
. See
the molbox Operation and Maintenance Manual, Ready/Not Ready Section, for
additional information.
molbloc
The serial number and molbox channel of the active molbloc. The molbloc channel
displays only when a two channel molbox is the active reference.
Содержание GFS2102
Страница 1: ...2007 DH Instruments a Fluke Company GFS2102 Gravimetric Flow Standard Operation and Maintenance Manual...
Страница 8: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 14: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XII N NO OT TE ES S...
Страница 16: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XIV N NO OT TE ES S...
Страница 20: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 4 N NO OT TE ES S...
Страница 32: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...
Страница 58: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 42 N NO OT TE ES S...
Страница 174: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 158 N NO OT TE ES S...
Страница 182: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 166 N NO OT TE ES S...
Страница 204: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 188 9 2 TYPICAL TEST SETUPS...
Страница 206: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 190 N NO OT TE ES S...