3.3.8
ENTER REFERENCE GAS CYLINDER DATA
Before making measurements, it is necessary to enter the technical data for each reference
gas cylinder assembly you have into GFS Tools software. This data can be found on the
technical data sheet that is delivered with each cylinder. Some of the data will be unique to
each cylinder so failing to enter the correct data can result in errors in the air buoyancy
correction that is made.
In GFS Tools, use the
[Setup][Reference Gas Cylinder]
menu choice to access the
Reference Cylinder Editor
. Fill in the Header Information and Cylinder Characteristics.
See Section 5.5.5 for details on the GFS Tools Reference Cylinder Editor. The Cylinder
Characteristics can be found on the Cylinder Technical Data sheet provided for each
cylinder.
3.3.9
INSTALLATION OF REFERENCE MASS ON MASS PLATE
Locate the reference mass and the AMH reference mass plate (the AMH plate with NO
bubble level on the top surface). The reference mass MUST be installed with the side etched
with the nominal mass and mass serial number facing AWAY from the AMH reference mass
plate so that these identifiers are plainly visible once installation is complete and the legs will
maintain the proper distance between the reference mass and AMH plate. The mounting
legs are indexed similar to the reference gas cylinder, two of the legs have a spherical curve
at one end and one has a flat. The legs with the spherical curve need to be mounted so they
correspond to the chamfer and groove brass inserts in the AMH plate. The leg with the flat
end corresponds to the flat brass AMH plate insert.
Figure 18.
Reference Mass Legs (Spherical (x 2) and Flat)
Invert the AMH plate painted side
down on a flat surface, taking care
not to damage the painted surface.
Padding or cloth of some kind would
help keep the surface from damage.
Step 1
Содержание GFS2102
Страница 1: ...2007 DH Instruments a Fluke Company GFS2102 Gravimetric Flow Standard Operation and Maintenance Manual...
Страница 8: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 14: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XII N NO OT TE ES S...
Страница 16: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XIV N NO OT TE ES S...
Страница 20: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 4 N NO OT TE ES S...
Страница 32: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...
Страница 58: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 42 N NO OT TE ES S...
Страница 174: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 158 N NO OT TE ES S...
Страница 182: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 166 N NO OT TE ES S...
Страница 204: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 188 9 2 TYPICAL TEST SETUPS...
Страница 206: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 190 N NO OT TE ES S...