
GFS2102™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 30
On the balance PRT tab, verify that the balance PRT calibration data matches the data
from the most recent LCM calibration report. This data is stored inside the LCM and is
read automatically into GFS Tools.
On the balance IR tab, verify that the infrared probe calibration data matches the data
from the most recent LCM calibration report. This data is stored inside the LCM and is
read automatically into GFS Tools.
3.3.7
ASSEMBLE REFERENCE GAS CYLINDER(S)
The GFS2102 reference gas cylinder assembly includes the reference gas cylinder, the
regulator assembly, the reference gas cylinder connector, and the cylinder support. (See
Figure 15.)
Figure 15.
Reference gas cylinder assembly
1.
Reference gas cylinder
2. Regulator assembly
3.
Reference gas cylinder connector
4.
Cylinder support
Never ship the reference gas cylinder assembly in the fill/carry case while the regulator
assembly is installed on the reference gas cylinder. This could lead to damage of the regulator
assembly, potentially causing a failure of the regulator assembly tubing under high pressure.
Energy stored in high pressure gases can be released unexpectedly and with extreme force.
High pressure systems should be assembled and operated only by personnel who have been
instructed in proper safety practices.
The reference gas cylinder assembly will arrive from the factory with the regulator assembly
disconnected from the rest of the cylinder assembly. All parts of the reference gas cylinder
assembly are shipped inside the carry/fill case. The regulator assembly is packed in a
custom foam cut-out inside the case to protect it from damage. The reference gas cylinder
Содержание GFS2102
Страница 1: ...2007 DH Instruments a Fluke Company GFS2102 Gravimetric Flow Standard Operation and Maintenance Manual...
Страница 8: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 14: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XII N NO OT TE ES S...
Страница 16: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XIV N NO OT TE ES S...
Страница 20: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 4 N NO OT TE ES S...
Страница 32: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...
Страница 58: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 42 N NO OT TE ES S...
Страница 174: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 158 N NO OT TE ES S...
Страница 182: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 166 N NO OT TE ES S...
Страница 204: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 188 9 2 TYPICAL TEST SETUPS...
Страница 206: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 190 N NO OT TE ES S...