GFS2102™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 8
2.2
REFERENCE GAS CYLINDER ASSEMBLY
The GFS2102 reference gas cylinder assembly is built around a high pressure carbon fiber composite
cylinder. In addition to the cylinder itself, the assembly includes a cylinder support, a cylinder connector
with fill port and valve, and a two stage welded regulator assembly.
Figure 5.
Reference gas cylinder assembly
1.
Reference gas cylinder
3.
Reference gas cylinder connector
2.
Regulator assembly
4.
Cylinder support
The reference gas cylinder is a metrological component of the GFS2102 system. Each cylinder assembly
is individually tested to determine the internal volume of the assembly. This information, along with the
external volume and coefficients of thermal and pressure expansion, allow proper calculation of an air
buoyancy correction on the measured mass value and the total mass of test gas available in the cylinder.
The characteristics of each cylinder are included on a technical data report that accompanies a new
cylinder assembly and must be entered in GFS Tools software (see Sections 3.3.8 and 5.7.5 for more
information).
The components of the reference gas cylinder assembly are designed or selected to be as light as
possible so that the mass of the test gas makes up a relatively large portion of total measured mass on
the balance, allowing sufficient resolution on the measurement of the mass change during the test. The
components of the GFS2102 reference gas cylinder assemblies are described below.
2.2.1
REFERENCE GAS CYLINDER
The cylinders used for GFS are composite cylinders made with a thin aluminum internal
lining, wrapped in carbon fiber in epoxy resin for strength and covered with a fiberglass lining
and gel coat finish for resistance to abrasion, impact and UV degradation. The fiber
composite design allows them to be lighter than most high pressure cylinders, so a smaller
percentage of the total filled cylinder assembly mass comes from the cylinder itself.
Содержание GFS2102
Страница 1: ...2007 DH Instruments a Fluke Company GFS2102 Gravimetric Flow Standard Operation and Maintenance Manual...
Страница 8: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 14: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XII N NO OT TE ES S...
Страница 16: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XIV N NO OT TE ES S...
Страница 20: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 4 N NO OT TE ES S...
Страница 32: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...
Страница 58: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 42 N NO OT TE ES S...
Страница 174: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 158 N NO OT TE ES S...
Страница 182: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 166 N NO OT TE ES S...
Страница 204: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 188 9 2 TYPICAL TEST SETUPS...
Страница 206: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 190 N NO OT TE ES S...