GFS2102™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 172
•
Adder is in the base unit of measure: Pascal (Pa) for pressure, °C for
temperature, %RH for humidity.
•
Multiplier is dimensionless
•
For the ambient PRT temperature, the GFS Tools
[Calibration]
function
reads and writes the resistance at 0 ºC (Rzero) and slope values of the PRT.
The Rzero and slope adjust the PRT as follows:
Corrected Output = (Measured Output – Rzero) / slope
Where:
•
Corrected output is in °C.
•
Measured output and Rzero are in ohms.
•
Slope is in ohms per °C, and is fixed at 0.3896 in its operating temperature
range (DIN Norm 43760). It is recommended that slope not be adjusted.
•
The calculation of a new Rzero is described in Section 7.5.4.3.
Within GFS Tools, the
[Setup]
,
[Calibration]
,
<Ambient>
tab provides
capabilities to edit the Adders and Multipliers of the ambient sensors and Rzero
and slope of the PRT. To calibrate or adjust the on-board sensors, compare
their outputs on the LCM front panel display to a reference and adjust the
corresponding Adder and/or Multiplier as needed to arrive at acceptable
agreement. Normally, only the Adder is used since a one point calibration is
being performed.
See Section 7.5.4.2 for information on calibration of the enclosure pressure
sensor.
See Section 7.5.4.3 for information on calibration of the ambient temperature
sensor.
See Section 7.5.4.4 for information on calibration of the ambient relative humidity
sensor.
1.
Current reading of the enclosure pressure sensor in
current pressure unit of measure.
2. Current reading of the ambient relative humidity
sensor in %RH.
3. Current reading of the ambient temperature PRT
in °C.
4.
Current reading of the Reference Mass external PRT
in °C.
97.003 kPa a 31.00%
22.032°C 23.536°C
Figure 110: LCM default front panel display
7.5.4.2
ENCLOSURE PRESSURE SENSOR
PURPOSE
To view and adjust the output of the enclosure pressure sensor.
OPERATION
The LCM default front panel display shows the output of the enclosure pressure
sensor in its upper left corner on power up (see Figure 110).
A one point verification against an atmospheric reference is all that is needed.
Содержание GFS2102
Страница 1: ...2007 DH Instruments a Fluke Company GFS2102 Gravimetric Flow Standard Operation and Maintenance Manual...
Страница 8: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 14: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XII N NO OT TE ES S...
Страница 16: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XIV N NO OT TE ES S...
Страница 20: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 4 N NO OT TE ES S...
Страница 32: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...
Страница 58: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 42 N NO OT TE ES S...
Страница 174: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 158 N NO OT TE ES S...
Страница 182: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 166 N NO OT TE ES S...
Страница 204: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 188 9 2 TYPICAL TEST SETUPS...
Страница 206: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 190 N NO OT TE ES S...