5. GFS TOOLS SOFTWARE
Page 97
© 2007 DH Instruments, a Fluke Company
The best fit DUT error is determined using the best fit order selected on the
[Best Fit]
tab. See
Figure 63
and
Table 35
for information on the options on the
[Best Fit]
tab. This is a convenient way of viewing predicted error information.
This option is not available for 3D plots and plots with a secondary Y axis. To
view best fit data when the
[Best Fit]
tab is not available, use the
Custom Plot
Editor
and specify best fit information for an axis in the plot.
Figure 62.
Plot Points display
Table 34.
Plot Points General Tab Fields
FEATURES
DESCRIPTION
Include tolerance
bars
(check box)
Causes tolerance bars to be displayed on the graph. The tolerance bars are calculated
using the tolerance value stored in the Data File. Tolerance bars are displayed only
when plotting errors and or flow information related to the DUT. Tolerance bars are also
not displayed when a secondary Y axis is included in the plot.
Limit Y axis to 1.5X
tolerance
(check box)
This option is active only when tolerance bars are included in the plot and an error plot
type is selected. If the error exceeds the tolerance by 50 %, the actual DUT error plot
will not be visible. When this function is not active, all plots are auto-scaled. This option
is often used for plots included in reports so that all plots are scaled in the same manner
regardless of actual error magnitude.
New subset for
each data point
(button)
Check this option to create a new subset of points for each flow point. Use this option to
generate lines of data for the same point when multiple cycles are used. For example, a
6 flow point 6 accumulation test would yield 6 lines of data. The error of each point can
be plotted to quickly display how error changed at a single point in each accumulation.
This feature is most useful when viewing accumulation information from the same flow
point.
Include subset
legend
(button)
Check this option to display a legend at the top of the plot to for each subset. Do not
check this option when many subsets will be used.
Содержание GFS2102
Страница 1: ...2007 DH Instruments a Fluke Company GFS2102 Gravimetric Flow Standard Operation and Maintenance Manual...
Страница 8: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 14: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XII N NO OT TE ES S...
Страница 16: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XIV N NO OT TE ES S...
Страница 20: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 4 N NO OT TE ES S...
Страница 32: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...
Страница 58: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 42 N NO OT TE ES S...
Страница 174: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 158 N NO OT TE ES S...
Страница 182: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 166 N NO OT TE ES S...
Страница 204: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 188 9 2 TYPICAL TEST SETUPS...
Страница 206: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 190 N NO OT TE ES S...