5. GFS TOOLS SOFTWARE
Page 89
© 2007 DH Instruments, a Fluke Company
Table 30.
Options, [Data Grid] Tab Selections
FEATURE
DESCRIPTION
Available fields
(list box)
Contains an alphabetically sorted list of the data field columns not currently included in the
Data
Grid Run Screen
. Individual fields can be dragged and dropped onto the
<Columns in grid>
list.
Alternatively, multiple fields can be selected and moved by using combinations of the shift, arrow
and control keys along with the shift right button
“>>>>>>”
.
Fields in grid
(list box)
These are the data field columns that display in the
Data Grid Run Screen
when a test is running.
The
<Point>
field is always included and is always at the top of the list. All other data fields can
be added, removed or arranged in any order. Double-click a field to remove it from the list.
Alternatively, multiple fields can be selected and moved by using combinations of the shift, arrow
and control keys along with the shift left button
“<<<<<”
. Move a field’s position in the list by
selecting the field then using the UP or DOWN arrow keys while holding the CTRL button.
Scroll to display
current data point
(check box)
When checked, the current data point is automatically displayed in the Data Grid run screen when
a test point is logged. If the grid is scrolled between points, and a new point is taken, the grid
automatically scrolls to the bottom to display the new point. The plot is also updated to force the
display of the current point in the current test cycle even if the plot has been customized to not
display the current cycle.
Uncheck this box to avoid automatic forced updating of the
Data Grid
and
Plot Run Screens
.
5.2.3
[UNIT OF MEASURE CONVERTER]
All unit conversions and custom units are handled by the
DHI
Unit of Measure Converter
.
Run the
Unit of Measure Converter
as a stand-alone application using the program group
icon, or use the
[Tools]
,
[Unit of Measure Converter]
option in
GFS Tools
. The
[Options]
menu in the
Unit of Measure Converter
contains features to allow custom unit creation.
Custom units created in this way are available throughout
GFS Tools
. Refer to the help
menu of the
Unit of Measure Converter
for details on unit conversion factors, supported
units and custom unit creation.
5.2.4
[REMOTE COMMUNICATIONS]
The
Direct Remote Communication
form is accessible by using the shortcut icon in the
GFS
Tools
program group or by using the
[Tools]
,
[Remote Communications]
menu choice in the
program. This tool provides a simple means of sending individual command strings to remote
devices and receiving and displaying the response.
Direct Remote Communications
is used
for discreet remote device communication and communications testing. Use this tool to
familiarize yourself with the device’s communications, for troubleshooting communications
problems and determining the appropriate commands to send prior to entering command
information into of the Remote Command Editor (see 5.5.3.4, Remote Command Editor). If a
command issued to a device does not give the desired result, refer to the device documentation
for information on how to troubleshoot the problem. Try cycling power on the device and also try
a command more than one time before concluding the command is in error.
DHI device COM2 options are not supported by the Direct Remote Communication form. To
test this functionality, manually set the device COM2 port to the RS232 settings of the device
connected to the COM2 port. Then send commands with the COM2 prefix required by the product.
In most cases, the prefix “#” will work. Refer to the Operation and Maintenance manual of the
specific product for details on the pass through support
Содержание GFS2102
Страница 1: ...2007 DH Instruments a Fluke Company GFS2102 Gravimetric Flow Standard Operation and Maintenance Manual...
Страница 8: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 14: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XII N NO OT TE ES S...
Страница 16: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XIV N NO OT TE ES S...
Страница 20: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 4 N NO OT TE ES S...
Страница 32: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...
Страница 58: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 42 N NO OT TE ES S...
Страница 174: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 158 N NO OT TE ES S...
Страница 182: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 166 N NO OT TE ES S...
Страница 204: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 188 9 2 TYPICAL TEST SETUPS...
Страница 206: ...GFS2102 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 190 N NO OT TE ES S...