USER Manual
49
8.3.2
Annual User Check
Users should review the following material yearly, and be sure to record their results
in the Maintenance Log Sheets located in Section 8.3.4.
①
Power button verification: If you press the power button for 1 second, a green LED
light will turn on and a beep will sound to signal that the machine has powered
on. While the machine is on, press the power button for 1 second to shut the
machine down. Please check after thirty seconds to verify that the machine has
indeed turned off.
②
Touch LCD verification: Touch each corner on the 4.3 inch LCD touch panel to
check whether the panel responds correctly. Also make sure under menu to set
the settings properly for kV, mA, and Exposure time.
③
Exp button verification: In the equipment menu, press LED once. The collimator
LED should turn on and the exp button’s blue LED should turn on. In this state,
press the Exp button again. The LED on the Exp button should turn yellow to
confirm the X-ray has been emitted.
④
Remote button verification: Connect the Remote button to ensure that the Exp
button does not operate. Now check to see whether the Remote button operates
like the Exp button.
⑤
Deadman button verification: Set the exposure time to 1.5 seconds. Then use the
Exp button to start the X-ray irradiation. Before the exposure time reaches 1.5
seconds, release the button to terminate the x-ray irradiation.