30
MAIN SYSTEM
140
TRION3-1810M-POWER-4
1
2
3
4
5
6
7
8
Replace the sub-module.
Secure the replaced sub-module with the torx screw (M2x4,
TX6). Max. torque: 0.2 Nm
Insert the TRION3-1810M-POWER-4 module into the housing
until a resistance appears.
Pull up the injector/ejector handle to latch the module. Tighten
the screws at the top and bottom of the TRION3-1810-
POWER-4 module front panel (4x) to secure the module.
Remove the TRION3-1810M-POWER module from the housing.
Loosen the torx screw (M2x4, TX6) which secures the sub-
module of the channel you want to replace.
Exchanging SUB-modules
Proper ESD precautions must be taken to avoid any damage to the unit.
Switch off the instrument and unplug all connected cables
including sensors from the TRION™ series modules.
Switch OFF
Loosen the screws at the top and bottom of the TRION3-
1810M-POWER-4 module front panel (4x) and pull down the
injector/ejector handle to release the module.