DS-CAL
SOFTWARE USER MANUAL
Illustration 26: Example of connecting 4xHS
‑
HV Sirius modules to DS
‑
CAL test board
Hint
For calibration of HV and HS-HV modules consider also the usage of proper multifunction
calibrator model for filter tests. See Table 3, in chapter Multifunction calibrator support.
Example of connecting 4xHS-HV Sirius modules to DS-CAL test board shows Illustration 26.
9.4.4. Sirius LV and HS-LV
Caution
NEVER connect LV or HS-LV module to the HV connector on the DS-CAL test board! If Sirius
UUT contains mixed modules (for example a combination of LV and HV modules with banana
fronts) make sure NOT to connect the LV or HS-LV modules to the HV connector on the DS-CAL
test board as this will incur permanent damage to the LV or HS-LV module! Always connect LV or
HS-LV modules according to the Table 13.
LV and HS-LV modules connect to the DS-CAL test board according to the Table 13.
Module (Connector)
DS-CAL test board channel (Connector)
LV (DSUB9), HS-LV (DSUB9)
DW-43/STG (DSUB9)
LV (BNC), HS-LV (BNC)
ACC (BNC)
LV (BANANA), HS-LV (BANANA)
ACC (BNC via BANANA-to-BNC adapter)
Table 13: Sirius LV and HS-LV connection to the DS-CAL test board
DS-CAL V20-1
38/67
Содержание DS-CAL
Страница 1: ...DS CAL SOFTWARE USER MANUAL DS CAL V20 1 1...
Страница 20: ...DS CAL SOFTWARE USER MANUAL Illustration 11 Users tab of the DSCalibrator DS CAL V20 1 20 67...
Страница 26: ...DS CAL SOFTWARE USER MANUAL Illustration 17 Prepare adjustment procedure window DS CAL V20 1 26 67...
Страница 55: ...DS CAL SOFTWARE USER MANUAL Illustration 46 Example of connecting the DSI 5A DS CAL V20 1 55 67...