DS-CAL
SOFTWARE USER MANUAL
8.5.1. Omission of tests
Special considerations are required when ommiting calibration procedure tests automatically provided
by the DSCalibrator software.
UUT
Result
SIRIUS-CHG,
SIRIUS-HS-CHG,
DSI-CHG-50
Any or all of Analog IN, SNR, Filters, TEDS and Charge tests fail on
adjustment/calibration procedure when Init/Prepare tests are
ommited.
KRYPTON 8xTH,
KRYPTON 16xTH
Analog IN tests fail on adjustment/calibration procedure when SNR
tests are ommited.
Table 8: Special considerations for ommision of calibration procedure tests
8.5.2. Different module grounding setting
DSCalibrator software will notify if groups of modules require different grounding setting by issuing a
warning before starting calibration procedure similar to as shown on the Illustration 20, page 43.
Important
Reported groups of modules must be calibrated separately! Additionally, only connections to
the modules being calibrated must be made to the DS-CAL test board and all others
disconnected!
Illustration 20: Warning issued by DSCalibrator when different module grounding setting is required
DS-CAL V20-1
30/67