![Dewert 2410XL/H Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/dewert/2410xl-h/2410xl-h_user-manual_2489879006.webp)
Dewert User Manual Model 2410XL - Updated 05.2021/1 MDR Page 6 of 12
Never operate the side guard if the hands, fingers, etc. of another person are located between the
bars or on the mechanism of the side guard. Risk of being crushed/pinched!!!
The laterally lowerable side guard is completely screwed in place. If the side guard becomes too loose or has
excessive lateral play over time due to use, it can be readjusted by tightening the screws. The moving parts of
the side guard should be re-lubricated slightly at regular intervals (spray oil, e.g. Ballistol).
Upholstery
The upholstery consists of a 1-sectioned mattress, but the head part can be moved separately. Hook-and-
loop tape on the back of the mattress allows it to be positioned and attached to the base plate of the table.
The core of the upholstery consists of a very high-quality and flexible cold foam material. Do not remove the
upholstery.
The base plate on the table frame is made of HPL (high pressure laminate). This material is characterised by
its high strength and resistance to ageing. It is resistant to disinfectants and can be easily cleaned due to its
closed surface.
Paper roll holder (optional)
The paper roll holder consists of a holding bar and retaining pins that are attached below the bumpers on the head
part. In addition to the stainless steel bar, the paper roll holding bar consists of a spring-loaded stainless steel
sleeve featuring a round steel gripping disc at the end. To insert the holding bar, feed the guide cotter pin of the
stainless steel bar into the rear hole of the retaining pin. Then push the sleeve with the gripping disc against the
spring force and feed the front guide cotter pin into the second hole. Then relieve the spring tension on the sleeve.
To release the paper roll holder, proceed in the same way.
3. Additional accessories (for user-specific table configuration), in extracts
•
Paper roll holder
•
DIN rail 350 mm
•
Fastening clamp for DIN rail
•
Infusion pole for fastening clamp
•
Storage basket
4. Technical data
4.1 Technical data for model 2410XL/H
Lying surface length (mm)
1,930
Lying surface width (mm)
650
Head part length (mm)
720
Head part adjustment range (gas spring)
0° / +70°
Upholstery height (mm)
100
Total length (mm)
2,030
Total width (mm)
800
Min.
– max. height (mm)
690
– 1,110
Weight (approx., depending on
equipment) kg
110
Max. patient weight (kg)
250
Technical data subject to change
The mechanics of the tables have been designed with extensive safety margins in mind. The max. patient weight
is provided with a 4-fold static safety factor, i.e. the design has been fully tested for a 4-fold load.