This table is not classed as surgical furniture and must, therefore, not be used for surgical purposes. 
The expected service life is 10 years or 100,000 drive cycles (double stroke = 1x up and down). 

 
 

1.5 Information on setup and use 

When packed, the device may be exposed to the following environmental conditions for approx. 3 months:   
     
Ambient temperature: -5° to +50° 
Relative humidity: 10% to 90% 
Air pressure: 500hPa to 1060hPa 
       
When transporting the device in a vehicle, it must be secured properly against moving. To do so, lock the 
castors (optional equipment) and ensure further safety measures. 
 
When setting up the device, do not lift it at the head part, as this may damage the head part and/or the 
release mechanism. 
Hold and lift the table on the left and right-hand side of the underframe. 
The table must stand securely on its feet or castors on a level, flat and solid surface. Before use, activate the 
brakes on the castors and make sure they are working properly.  
 

When transporting the table, take hold of the underframe, NOT the upper frame. 

 

 

1.6 Commissioning 

The device is ready for use upon delivery. Lock the castors (optional). As the operator, carry out a thorough 
and precise function check once the device has been set up. Prior to commissioning, clean the device and 
remove any contamination from transport. Operation in potentially explosive atmospheres is not permitted.  
 
 

1.7 Safety instructions  

This table may only be used for its intended purpose. Any other use is strictly prohibited and possibly 
dangerous. The manufacturer cannot be held responsible for damage caused by improper use. 
Patients may only be positioned in preparation for treatment/examination by professionally trained persons. 

Please note:

 This table is not classed as surgical furniture. 

 

Prior to and when adjusting the height of the table, make sure that no persons or objects are located in the 
adjustment range of the table and that nobody has their hands on the underframe

.

 

 

The following basically applies: Never reach into or under the frame of the table when adjusting the height. 
Height adjustment can result in injury if the user does not pay due care and attention. Therefore, take great 
care when performing this procedure. 

 

When adjusting the upholstery parts, make sure that no persons or objects are in the adjustment range. Make 
sure that no persons reach under the upholstery part or lean on the underframe. 
 
Important for the user: When adjusting the upholstery parts, do not reach under the spacers located beneath 
the upholstery parts. 

 

Always use both hands when adjusting the lying surface elements: Use one hand to operate the adjustment 
mechanism and the other hand for the lying surface adjustment. 

 

The lying surfaces and the underframe are not anti-static as standard. 
Our products are not intended for use in wet rooms and must under no circumstances be cleaned using so-
called bed washers. This would irreparably destroy the product. 

 

The head part is only intended to support the patient and must not be used for sitting. 

 

The wheels must always be locked before using the table.  

 

Do not put a damaged device into operation. 
 
This device must not be modified without the express permission of the manufacturer.

 

Содержание 2410XL/H

Страница 1: ...Safety regulations 2 1 4 Intended purpose 2 1 5 Information on setup and use 3 1 6 Commissioning 3 1 7 Safety instructions 3 1 8 Model designation and type labelling 4 1 9 Meaning of the serial number...

Страница 2: ...g manual fully The manufacturer cannot be held responsible for damage caused by or involving unauthorised persons The user must ensure that the device is not accessible to unauthorised persons or cann...

Страница 3: ...strictly prohibited and possibly dangerous The manufacturer cannot be held responsible for damage caused by improper use Patients may only be positioned in preparation for treatment examination by pr...

Страница 4: ...al safety The number of possible pinching points has thus been minimised while remaining ones have been covered or protected with spacers to prevent injury thus ensuring safe and yet simple operation...

Страница 5: ...with infusions 2 4 Movability The table is equipped with easy running comfort castors measuring 150 mm in diameter in a conductive design Central movability By operating a lever on the outside of the...

Страница 6: ...ts of a holding bar and retaining pins that are attached below the bumpers on the head part In addition to the stainless steel bar the paper roll holding bar consists of a spring loaded stainless stee...

Страница 7: ...vents dust from entering through the dust lip and preserves the service life of the unit Caution When cleaning secure the table against unintentional lowering of the lying surface To do so set all the...

Страница 8: ...liable for damages resulting from uncoordinated modifications Only original Dewert spare parts may be used Gas springs are wearing parts whose function may be impaired over the years depending on the...

Страница 9: ...sposal costs meaning you can dispose of the packaging free of charge Product Safety notices o Pay attention to sharp edges and pointed objects o When transporting the table only take hold of the under...

Страница 10: ...similar medical devices have knowledge of its proper use taking into account possible hazards Ideal positioning of patients for the purpose of transport to pre treatment or post treatment locations a...

Страница 11: ...damaged All switches and supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off...

Страница 12: ...ssories Accessories such as belts belt pads belt guides paper roll holders armrests etc undamaged and is proper and secure fixing function possible All necessary toggle screws present Possible movabil...

Отзывы: