Desco 19770 Скачать руководство пользователя страница 1

TB-2024

11/01 Page 1 of 4

Digital Surface Resistance/Resistivity 
Test Kit Operation and Maintenance

Figure 1. Desco Item 19770

Description

The LCD Surface Resistance/Resistivity Test Kit is a
portable instrument designed to measure resistance
between two points (RTT), surface to ground (RTG), and
surface resistivity in accordance with ANSI EOS/ESD
Association Standard S4.1. The unit’s unique Liquid Crystal
Display (LCD) allows the user to easily read measurements
directly from the meter. No interpretation or calculations are
necessary. Per ANSI/ESD S20.20, test equipment is
required to enact Compliance Verification
Plan Requirement - Test equipment shall be selected to
make measurements of appropriate properties of the
technical requirements that are incorporated into the ESD
program plan. This unit is ideally suited for evaluating the
electrical properties of ESD protective work surfaces and
flooring products.

The LCD Surface Resistance/Resistivity Test Kit is NIST
calibrated and is available in three models:

Figure 2. Parallel Resistivity Electrodes on bottom of meter.

The meter also incorporates two parallel probes on the
bottom of the unit which allows surface resistivity
measurements. These parallel electrodes allow for quick
and easy testing of a variety of surfaces and materials,
without the use of 5 pound electrodes. The 5-pound
electrodes (item 50003), and test leads (item 50075) can be
bought separately.

The Test Kit displays a measurement per ESD S4.1-1997
proving the proper electrification period of 15 seconds
during which numerous readings and calculations are
executed. 

NOTE: 

Testing in accordance with ESD S4.1-

1997 requires 15 seconds of electrification; in contrast, most
analog type meters display measurements instantaneously.

Recommended Literature

Desco recommends that you read the following standards
from the ESD Association:

ANSI/ESD S20.20 — Development of ESD Control Program
ESD ADV1.0 — Glossary of Terms
ESD S4.1 — Worksurfaces
ANSI EOS/ESD S6.1— Grounding
ANSI ESD S7.1 — Floor Materials
ANSI EOS/ESD S11.11 — Surface Resistivity

These documents can be obtained directly from the ESD
Association, 7902 Turin Rd., Suite 4, Rome, NY 13440-
2069, (315) 339-6937.

In addition to the Association Standards listed previously,
anyone testing the electrical properties of ESD protective
surfaces should also obtain copies of:

MIL-HDBK 263A

EIA-IS-5-A

ASTM-F-150

EN 100015

EIA-625

These standards are available from the agencies who
produce them. If you need help in obtaining these
documents, please contact our customer service
department.

Model

Description

Voltage/Hz 

19770 Test 

Kit

120/50-60

19775 Test 

Kit

220/50-60

19771

Tester Only

120/50-60

DESCO WEST

- 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449 

DESCO EAST

- 90 Hudson Road, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: http://www.desco.com

TECHNICAL BULLETIN TB-2024

Made in America

Содержание 19770

Страница 1: ...5 pound electrodes item 50003 and test leads item 50075 can be bought separately The Test Kit displays a measurement per ESD S4 1 1997 proving the proper electrification period of 15 seconds during which numerous readings and calculations are executed NOTE Testing in accordance with ESD S4 1 1997 requires 15 seconds of electrification in contrast most analog type meters display measurements instan...

Страница 2: ...he measured reading will remain illuminated for at least 10 seconds C Test Range Voltage Switch 10 Volts for 1x103 1x1011 ohms 100 Volts for 1x106 1x1012 ohms D Jacks 2 5mm plugs will fit the megohmmeter jacks Banana plugs will fit the 5 lb electrodes E AC Battery Charger Jack 12 volts DC 100mA F 9 Volt Battery Compartment G Parallel Electrodes Figure 3 Features of the LCD Surface Resistance Resis...

Страница 3: ...re selected Set the meter to the required test voltage determined by the test procedure and the resistance of the material Conductive material 104 ohms or less is usually tested at 10 volts Dissipative material is usually tested at 100 volts If material to be tested is unknown it is recommended to first test at 10 volts C Press and hold the red test button for 10 20 seconds the unit will be making...

Страница 4: ... the resistors to the enclosed cords using grounding clips and recording the meter display Keep the cords separated Adjustments to the internal resistance pots can be done quite easily Request this procedure from the Desco Customer Service Department Specifications Ranges 103 1011 ohms 10 Volts 106 1012 ohms 100 Volts Power Supply 9 volt rechargeable Nickel Cadmium battery and AC adapter charger i...