www.StylusATC.com

Technical Customer Service / Air Repair

1.800.800.7202

Code No. 87135-0610

©2010 DENTSPLY. All rights reserved.

The Midwest

®

 Stylus

 ATC

Handpiece Maintenance Instructions

Refer to DFU included with handpiece for detailed instructions on operation,  
sterilization, warranty and ordering Information. Additional information can be  
obtained at 

www.stylusatc.com

 or by calling 

1.800.800.7202

After Every Use

1

2

3

4

5

6

7

8

Disconnect handpiece from coupler.

Press push button and remove bur.

Connect ATC Lubrication Nozzle and Cap 
assembly onto Midwest Aerosol Spray.

Fully insert nozzle into handpiece, 
listening for a “click.” 

Cover handpiece head with paper  
towel. Hold spray can upright; spray  
for one second.

Connect ATC Maintenance Coupler  
to back end of handpiece. 

Cover ATC handpiece head with  
spray guard. 

11

Wipe off excess lubricant on outside  
of handpiece with dry gauze pad or  
paper towel.

12

Steam sterilize handpiece (without bur)  
at maximum 135˚C/275˚F (bagged cycle).

10

If not using air station, connect to  
4- or 5-hole hose to expel. 

Caution:

 ATC Maintenance Coupler 

must be used to expel lubricant. Do  
not use the ATC controller unit to expel.

Scrub and rinse handpiece under warm 
water with soft sponge. Scrub fiber-optic 
light port well.

9

Connect handpiece to Midwest air  
station with ATC Maintenance Coupler. 
Run air through handpiece for 30 
seconds at 40 psi to expel lubricant.

ATC Lubrication Nozzle Capp (880080)

Midwest Aerosol Spray (380080)

ATC Maintenance Coupler (880090)

Spray Gaurd (380150)

Midwest Air Station (380800)

Spray Guard (380150)