7 Maintenance and cleaning
Operating Instructions inLab MC X5
7.9 Procedure when processing Medentika-PreFace® abutment semi-finished products made of
båÖäáëÜ
7.9
Procedure when processing Medentika-
PreFace® abutment semi-finished products
made of titanium
7.9.1
Important information
CAUTION
Risk of injury
There is risk of injury through sharp-edged restorations as well as sharp-
edged material residues in the tank.
➢ Pay attention to the sharp-edged restorations / material residues
after processing, during cleaning and coolant changes.
NOTICE
Observe the safety information from the material manufacturer
Observe the safety instructions regarding occupational safety and
disposal referred to in the material manufacturer's operating
instructions. Dirty filters, tanks and metal residue must be disposed of
appropriately in accordance with these specifications.
IMPORTANT
When processing workpieces from the material class metal (Medentika-
PreFace
®
abutment semi-finished products), use the "Starter Kit
Preface Abutments Medentika" (REF 6597210).
It contains all of the relevant equipment required to process workpieces
(holders, tools, coolant lubricant additive DentaLub, tool magazine and
two coolant tanks. Use one of the tanks for processing and the other to
clean the system.)
IMPORTANT
Use a coolant-lubricant mix of water and the coolant lubricant additive
DentaLub to process titanium workpieces.
IMPORTANT
After all processing using the coolant lubricant additive DentaLub before
switching to another material, perform a cleaning process with then
intended cleaning tank (see "After processing / before changing to
processing another material [ → 97]").
IMPORTANT
Ask your disposal company whether the filter waste needs to be
correctly sorted for disposal.