7 Maintenance and cleaning
Dentsply Sirona
7.7 Coolant system
Operating Instructions inLab MC X5
7.7.3
Replace the coolant when using DENTATEC (for material
classes PMMA/wax, composite and grinding materials)
We recommend changing the coolant after one month at the latest.
7.7.3.1
Preventing odors
All coolant additives contain a biodegradable preservative. Despite this,
however, odors may still develop under unfavorable conditions.
Observe the following:
● Change the coolant at least once a month.
● Drain the tank if you do not intend to operate the unit for more than
one week.
● Clean the tank if the odors recur.
● Add the coolant additive DENTATEC and fill the tank up to the brim
with water. Let it stand for at least 24 hours and then rinse it out
thoroughly with water once again.
NOTICE
Damage to the pump and the tool mandrel!
Too high a proportion of ceramic in the coolant will damage the pump
and the tool mandrel.
Change the coolant regularly!
NOTICE
Damage to surfaces!
In the undiluted state, the coolant additive DENTATEC disintegrates
plastic surfaces and can cause discoloration.
➢ Do not place DENTATEC on the unit.
➢ Do not spill DENTATEC.
NOTICE
Approved coolant additive
Use only DENTATEC as a coolant additive.