7 Maintenance and cleaning
Operating Instructions inLab MC X5
7.11 Error messages and measures
båÖäáëÜ
7.11
Error messages and measures
In the event of an error, the machine outputs error messages. Refer to the
following table for measures to be taken for the respective error.
Error message
Action
Error in the air circuit of the spindle sealing
air.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Error in the air circuit of the tool change
cylinder.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Error in the blast air circuit.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Error in the air circuit of the boost valve
(pressure increase).
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
The water pump reported an error.
Please check that there is enough water in the tank or that the inlet and
outlet are properly connected.
The suction device reported an error.
Please check that the suction device is connected and switched on or,
for wet processing, is switched off. Please also check that the plugs are
properly connected.
The spindle reported an error.
Please check the state of the tools. If the tools are heavily worn out,
replace them and re-start the process.
Internal software error.
Please re-start the machine. If the error persists, contact customer
service.
No system pressure was detected.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Unspecified communication error. Please
try again.
Unspecified communication error. Please try again. If the error persists,
contact customer service.
Time-out occurred during order processing
!
Unspecified communication error. Please try again. If the error persists,
contact customer service.
The suction device reported an error.
Please check that the suction device is connected and switched on and
check that the plug is properly connected.
The suction device reported an error.
Please check that the suction device is switched off.
No valid calibration data available.
Please contact customer service.
The door lock reported an error.
Please check that the door is properly locked. If the error persists,
contact customer service.
The cover of the instrument tray cannot be
closed.
Please check that the lock of the magazine is closed.
The chuck cannot be opened.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
The chuck cannot be closed.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
No tool detected.
Please check that each tool is connected correctly.
The tool could not be removed and is still
secure in the chuck.
Please try again to change the tool or change the tool manually using
the spindle maintenance in the machine configuration. Clean the chuck
using the accompanying maintenance kit.
The tool sensor reported an error.
Please clean the sensor and then manually actuate it a few times. If the
error persists, contact customer service.