Step by Step Instructions
26/74
BENPPIQDFUWEB / Rev.04 / 08-2018
6.8.2
Plugging in the X-Smart IQ
®
Measurement Cable in
combined mode
N°
Action
Only use
X-Smart IQ
®
Handpiece for working in combined mode. Using
Propex IQ
®
Apex Locator with another Handpiece can lead to severe electrical problems (including
electrical shock).
A
Insert
X-Smart IQ
®
Handpiece inside its proprietary single-use protective sleeve.
The use of the
X-Smart IQ
®
Handpiece Sleeve is mandatory to avoid contact between
the contra-angle and the patient’s mouth, which might cause wrong readings on
Propex IQ
®
Apex Locator.
B
Insert
Propex IQ
®
Apex Locator inside its proprietary single-use protective sleeve.
C
Plug in the
X-Smart IQ
®
Measurement Cable (
2
) to
Propex IQ
®
Apex Locator (
1
).
D
Connect the
Propex IQ
®
Apex Locator Lip Clip (
3
) to the
X-Smart IQ
®
Measurement
Cable (
2
).
E
Plug in
X-Smart IQ
®
Handpiece (
4
) to the
X-Smart IQ
®
Measurement Cable (
2
).
For further information on
X-Smart IQ
®
Handpiece Sleeve, please refer to
X-Smart IQ
®
Handpiece Directions For Use.
1
2
3
4
The
X-Smart IQ
®
Handpiece sleeve
supplied in your country may look
different from the
X-Smart IQ
®
Handpiece sleeve shown on the
illustration.