background image

16

4.3 Self-test

Each time the unit is turned on, the self-test program is executed to check important functions.  

The following functions will be checked:

Lift (firing chamber)

Vacuum

Pressing function ( Multimat® Cube press only)

Temperature

 

Touch <OK> to start the self-test immediately.

If a touch display field is not touched within  

30 seconds, the self-test will start automatically. 

 

Touch <Cancel> if no self-test is required.

If the self-test has been passed:

 

Touch <Commit> to complete the self-test.

If the self-test is not passed, the faulty function is 

crossed out in red.

 

Touch <Commit> to complete the self-test.

For possible problems and remedies:  

Section 5 “Troubleshooting”.

Содержание Multimat Cube

Страница 1: ...Instruction Manual Multimat Cube Multimat Cube press...

Страница 2: ...the unit 8 2 3 Accessories optional 8 2 4 Technical specifications 9 3 Unpacking and commissioning 10 3 1 Unpacking the unit 10 3 2 Inspecting the shipment 10 3 3 Connecting the unit 11 4 Operation 12...

Страница 3: ...34 35 6 3 Updating the firmware and software 36 7 Waste disposal 37 7 1 Disposing of the packaging 37 7 2 Disposing of the unit 37 8 Warranty 38 Copyright This instruction manual is protected by copy...

Страница 4: ...anted by Dentsply Sirona Examples of conceivable improper uses Operation without a firing base or pressing base Multimat Cube press only Firing or pressing Multimat Cube press only of unapproved mater...

Страница 5: ...ng to the requirements laid down in the instruction manual Compliance with all warnings and safety instructions Compliance with service instructions shown on the unit s display Service and repair work...

Страница 6: ...shing and impact hazard due to falling loads The unit may cause injury if dropped Always have two people lift the device together Transport the unit in its original packaging CAUTION Damage due to imp...

Страница 7: ...9 Pressing chamber 10 USB port for WLAN stick 11 Connector for compressed air connection hose 12 Connector for vacuum hose 13 LAN connector 14 IEC cold appliance socket for vacuum pump 15 Feet of the...

Страница 8: ...ight signals green ready flashing green program finished flashing red error blue unit in operation flashing blue action required pulsating grey white standby mode 2 3 Accessories optional Vacuum pump...

Страница 9: ...Ambient temperature operation 5 C to 40 C ideal 20 C Ambient temperature storage 0 C to 40 C Humidity 80 relative humidity at up to 31 C or linear decrease to 50 at 40 C Environment Indoors only on a...

Страница 10: ...packaging WARNING The magnetic firing supports pose a hazard to wearers of implanted devices Keep the magnetic firing support at least 10 cm away from the upper body 3 2 Inspecting the shipment Scope...

Страница 11: ...for the vacuum pump Multimat Cube press only Push the compressed air connection hose into the clamping coupler of the filter pressure regulator 2 The compressed air connection hose is sufficiently se...

Страница 12: ...on the screen until the desired selection aperes in the centre Scroll bar Scroll the screen up or down Use the scroll bar on the right side of the screen or window to move the screen contents up or do...

Страница 13: ...ssing base Multimat Cube press only Place the firing base or pressing base in the middle of the firing base carrier 1 Multimat Cube press only Attach the magnetic firing supports to the unit 2 Touch t...

Страница 14: ...ivating the screen saver Setting the time to 00 00 will deactivate the screen saver Changing the unit for temperature display Touch Temperature unit and select Celsius C or Fahrenheit F Slide mode sel...

Страница 15: ...tion the pressing programs that are available during operation can be preselected hiding unneeded firing programs Remove the tick mark near a program if it is not needed Touch Commit Vorauswahl der Br...

Страница 16: ...ure Touch OK to start the self test immediately If a touch display field is not touched within 30 seconds the self test will start automatically Touch Cancel if no self test is required If the self te...

Страница 17: ...d to check the pressing pressure If the tolerance limit is exceeded or not reached during the execution of a pressing program a message will be displayed after the end of the program The pressing proc...

Страница 18: ...r fast action 1 Status bar Current menu 2 Quick selection field Firing 3 Quick selection field Pressing 4 Quick selection field Last 10 programs 5 Status bar Pressure and temperature display 6 Main me...

Страница 19: ...19 4 5 2 Main menu 4 5 3 Settings menu Touch Settings 1 Home field 2 Firing menu 3 Pressing menu 4 File Manager menu 5 Settings menu 6 Standby mode field 7 Fast cooling Multimat Cube press only...

Страница 20: ...me Return furnace to factory setting Furnace menu Base temperature Standby temperature Calibrating Preheating program ON OFF Start preheating program Vacuum test Information menu Software version Firm...

Страница 21: ...fast cooling Touch The firing chamber will be lifted and the vacuum pump will run until the base temperature is reached 4 5 6 File Manager menu General notes Predefined programs cannot be edited or d...

Страница 22: ...of screen action page To select a group or program touch a group twice in succession to access the subgroup subfolder This subgroup may contain other subgroups or programs To highlight groups or progr...

Страница 23: ...to be created Touch the Firing programs field 1 Touch My Programs twice 2 to access the subgroup Touch the Create new activity 3 in the right hand window Touch Group 1 Touch Name 2 A keyboard will be...

Страница 24: ...Create a new program Touch the Firing programs field 1 Switch to the subfolder Touch My Programs twice 1 to access the subgroups Touch Test Group twice to access the Testgruppe subgroup Touch the Cre...

Страница 25: ...Test 1 1 Touch the Edit activity 3 in the right hand window Edit program Test 1 Touch the various fields 1 to modify the firing parameters Saving modified parameters Touch the Save field 2 Discarding...

Страница 26: ...access the subgroups Touch Oxide Firings twice 2 to access the subgroup with the firing programs Touch the Copy activity 3 in the right window Change to the Test Group My Programs subgroup Touch My Pr...

Страница 27: ...the Favourites selection on the Home screen Touch Home Deleting a favourite Select the favourite program or group Touch the Favourites field The program or group will no longer be marked with an aste...

Страница 28: ...chamber closed Vakuum Specifies the vacuum level in the firing chamber Firing stages Specifies the number max 3 of stages holding levels used Heating rate Specifies the speed of the temperature rise...

Страница 29: ...nshots were taken with slide mode ON Menu Settings General Slide mode Place the object to be fired in the centre of the firing base or pressing base Multimat Cube press only Swipe to the left or right...

Страница 30: ...r starting the program provided the pro gram is not running yet Parameters in predefined programs can only be changed for the current procedure the changes cannot be saved To save parameter changes pe...

Страница 31: ...the unit is turned on you will be prompted to run the pre heating program If the pre heating program was run completely and without aborting and the unit was subsequently turned off the next prompt t...

Страница 32: ...extended during the first 15 seconds of the pressing phase Check the pressing ring Pressing system error Pressing plunger extended too far Program must be aborted Pressing plunger completely ex tende...

Страница 33: ...he program was completed Supply voltage failed briefly while the program was executed Check the firing pressing result Battery error 1 The battery of the control unit is empty Contact Dentsply Sirona...

Страница 34: ...a prompt will appear asking whether the calibration value should be accepted Touch No Set Preheating time to 03 00 minutes VORSICHT Burns hazards when touching hot surfaces Allow the unit to cool befo...

Страница 35: ...oo low The silver wire has almost or completely melted No melting effect is visible on the silver wire Bewertung Weiteres Vorgehen Temperature correct Touch Yes The calibration value is automatically...

Страница 36: ...ona into the USB port on the left side of the unit Touch Settings menu General Update New versions are displayed Touch Install Version XXXXX A message window displays the new functions Touch Install n...

Страница 37: ...aw materials The external packaging is made of cardboard The filler materials and insets are made cardboard or polyethylene foam The protective foils and bags are made of polyethylene PE Dispose of th...

Страница 38: ...Damage caused by improper repairs and modifications made by unauthorized persons This warranty does not cover any further claims including but not limited to material and immaterial claims This device...

Страница 39: ...39...

Страница 40: ...50572601 REV 2019 03 EN Made in Germany Manufacturer DeguDent GmbH Rodenbacher Chaussee 4 63457 Hanau Wolfgang Germany Telefon 49 61 81 59 50 www dentsplysirona com...

Отзывы: