6 Post-reprocessing
Operating instructions
34
66 27 587 D3673.201.02.04.02
6
Post-reprocessing
6.1
After each treatment session
NOTE! Condition immediately, or at the latest, one hour after
treatment.
NOTE! Use only Midwest Plus Spray.
ü
The bur must not be moving.
ü
Wear appropriate protective clothing.
1.
Purge the water and air channels on the treatment center for
30 seconds.
2.
Remove the bur with tweezers.
3.
Predisinfect the high-speed handpiece and the quick coupling
directly at the treatment center.
4.
Take the high-speed handpiece and unscrew the quick
coupling.