Dentsply Sirona
Operating Instructions DAC Universal
5 Cleaning, maintenance and validation
5.1 Check & Clean process
6681006 D3708
D3708.201.01.01.02 04.2019
85
5.1.1
Start Check & Clean
ü
The DAC Universal is switched on and cooled down.
➢ Start the Check & Clean process.
The Check & Clean process can be started by either pressing
● the "
Check & Clean 50" function key
● the "
Check & Clean 200" function key
directly on the start dialog (if active) or via the menu with the function
key.
The Check & Clean process can be interrupted with pop-up messages.
The "
Check & Clean not possible" function key appears on the display.
Rectify this error to start the Check & Clean process.
5.1.2
Put on gloves
Frequency: Weekly or every 50 cycles (whichever comes first).
04:00 PM
For your own protection, put on gloves before you start the Check &
Clean process.