3 Product description
3.4 Technical data
Dentsply Sirona
Operating Instructions DAC Universal
28
6681006 D3708
D3708.201.01.01.02 04.2019
3.4
Technical data
Manufacturer:
Sirona Dental Systems GmbH
Model:
DAC Universal
Dimensions (H/W/D):
Opened: 59 cm / 40 cm / 42 cm
Closed with Blue lid: 40 cm / 40 cm / 42 cm
Materials:
Plastic, stainless steel
Weight: (net without packaging)
26 kg / 57 pounds
Weight (gross with packaging and accessories):
34 kg / 75 pounds
Noise:
< 65 dBA
Power supply:
100-127 VAC / 220-240 VAC
Frequency:
50 Hz / 60 Hz
Nominal current:
120 / 230 VAC: 11 A / 6 A
Fluctuations in electrical supply
+/- 10 % of nominal voltage
Transient overvoltage
Installation category II
Power consumption
1.3 kW
Protection class:
I
Protection type (IP class):
21B
Type:
Not applicable
Type BF applied part:
Not applicable
Safety class:
Never use the DAC Universal in the presence of
flammable anesthetics or gases.
Cycle length:
approx. 15 minutes
Operating conditions:
Temperature: 16 °C - 30 °C (61 °F - 86 °F)
No operation above 2000 m.
Relative humidity
up to 80%
Air pollution level (ISO 8573-1:2010)
≤ 2
Maximum loss of heat at a surrounding temperature of
approximate 23 °C (73 °F)
about 500 W for the whole device
about 250 W for the front.
Total chamber volume
< 2 liters
Design pressure in the chamber
3.5 bar relative
Maximum allowable working pressure in the chamber
3.5 bar relative
Maximum allowable working temperature in the cham-
ber
138 °C (280 °F)
Usable space in the chamber
Height: Approximate 16 cm,
Diameter: Approximate 11 cm
Transport and storage conditions:
Temperature: 10 °C - 40 °C (50 °F - 104 °F)
Relative humidity: 30 % - 95 %
Air pressure: 700 hPA - 1,060 hPa