8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
66 37 487 D3652
D3652.201.01.21.02 03.2021
65
English
8.1.2.9
High-level disinfection of the mirror sleeve (without HLD Set)
In case the HLD Set is not available
NOTE
Dry heat sterilization and high-level disinfection must not be
combined.
1.
Pre-cleaning (see „Pre-cleaning of the mirror sleeve [→ 59]“).
2.
Remove the mirror sleeve from the CEREC Omnicam (see
„Removing the mirror sleeve [→ 58]“).
3.
Place the camera in its camera cradle.
4.
Use the following disinfectant for the high-level disinfection: CIDEX
®
OPA.
CAUTION
Observe the disinfectant‘s manufacturer‘s safety indications!
NOTE
Only place the mirror sleeve into the container and not the entire
CEREC Omnicam.
Ensure that the mirror sleeve is placed upright in the disinfectant (with
the coated sapphire glass down) so that the solution cannot penetrate
the inside of the mirror sleeve.
5.
Choose a container. Find an example on the left.