Motor Handpiece Error Code
F1902142.EN / 05 / 2015 / updated 02/2016
65/74
9
Motor Handpiece Error Code
If a failure occurs on the motor handpiece, the nature of the failure is
given by a combination of colors as described in the following table.
State
Failure
Solution
Total system
failure
Re-initialize the motor handpiece with a
Factory Reset (see
Battery empty
Charge the battery (see
).
Motor
overheating
Wait for a few minutes and press the
multifunction button to reactivate the motor
handpiece.
Motor
hardware
failure
Reset the device (see
).
Battery failure
Return the instrument to your service center.
General
hardware
failure
Reset the device (see
).
OFF