866.DTE.INFO (383.4636)
www.DentalEZ.com
0344
StarDental®
1816 Colonial Village Lane
Lancaster, PA 17601 U.S.A.
Printed in the U.S.A.
©2019 DentalEZ, Inc.
Emergo Europe
Prinsessegracht 20
2514 AP The Hague
The Netherlands
EC REP
StarDental®
Instruments
Caution:
Federal law restricts this device to
sale by or on the order of a dentist.
065810 E Date of Issue 2019-05-24
Titan®-E Low-Speed
Handpiece Motor
Instruction Manual
Non-Sterile
Medical Device
Contents
Specifications
20,000 rpm Motor:
StarDental
®
Part #266131 (Fixed Back End Part #266132)
• No Lubrication Required
• Recommended Air Pressure: 40-45 psi (swivel motor)
• Air Flow: 2.1 SCFM @ 40-45 psi at Maximum Power Point (swivel motor)
• Rotation: Forward and Reverse
• 4-Line Connection (Fixed Back End)
Installation
1. Select the motor appropriate for tubing based on power requirements. Examine connection
of the motor and swivel tubing configuration as these must match configuration of delivery
system tubing.
Drive Air
Water
Exhaust
Air
Air
4-Line Tubing
2/3-Line Tubing
Drive
Air
Water
• Specifications
•
Installation
• Operation
• Cleaning & Sterilization
• Troubleshooting
• Replacement Part Numbers
• Warranty
Sterilizable up to 135°C
135°C
Titan E
20,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
®
StarDental
®
20K Motor
(
StarDental
Part #266131)
Swivel
BEFORE OPERATING THE MOTOR, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS. SAVE
THESE INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL WARNINGS AND CAUTIONS.
Indications For Use
Used intraorally by trained dental professionals for drilling and preparation of dental cavities
for restoration, such as fillings. The intended use extends to disking, cavity and crown preps,
polishing, post and pin drilling and pin setting.
Swivel
Bearing Assembly
ISO Connector
Tubing
Coupler
Tubing
2. Disconnect swivel from handpiece if necessary. Grasp the swivel in front of the bearing
assembly between the thumb and index finger as shown.
3. Slide tubing coupler until the end of the ISO connector is exposed; hold in hand as shown
with fingers securing the tubing.
4. Align and insert swivel tubes into the ISO connector holes.
5. While applying slight pressure to hold fittings in the connector holes, slide coupler forward
and engage threads. Continue threading coupler clockwise until hand tight against the swivel.
Wrench
Tubing
Coupler
Turn Coupler While Holding
Swivel in Place
6. Using the wrench as shown, tighten the swivel to the tubing coupler. Be sure to turn tubing
coupler, not swivel. DO NOT OVER TIGHTEN.
Note:
This joint loosens over time, resulting in loss of power. Check for tightness periodically.
7. Connect handpiece to swivel.
8. Using an air pressure gauge for 4-line,
StarDental
Part #262591, between swivel and tubing,
obtain a reading. Adjust the air pressure at delivery head until gauge reads 40-45 psi while
operating handpiece in full forward position and with foot control completely depressed.
Note:
If delivery unit has oil/mist lubrication, detach and drain; handpiece needs no lubrication.
Installation - Fixed Back End Motor
Titan E
20,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
®
StarDental
®
Direction & Speed
Control Ring
Rear End Connector
20K Fixed Back End Motor
(
StarDental
®
Part #266132)
Note
: If delivery unit has oil/mist lubrication, detach and drain; motor is LUBE FREE.
NOTICE
To prevent damage to bearings, before installing always check drive air pressure by
using air pressure gauge,
StarDental
Part #262591 (4/6-Line) between motor and tubing and
obtain a reading. Adjust air pressure at delivery head until gauge reads 40-45 psi while operating
motor and with foot control completely depressed.
1. Examine tubing. Verify tubing uses 4-Line connection.
2. To connect motor to tubing, align drive air and exhaust tubes with mating ISO connector holes
in coupler.
3. While applying slight pressure to hold tubing in connector holes, slide tubing coupler forward
and engage threads. Continue threading tubing coupler clockwise until
hand tight against the motor. DO NOT OVER TIGHTEN.
Note
: This joint loosens over time, resulting in air leakage and loss of power. Check for
tightness periodically.
Operation
The 120° rotation speed adjustment ring sets direction and speed. To run motor forward at full
speed, rotate ring towards the letter ‘F’ while operating the foot control. Rotate to the letter ‘R’ to
yield full speed in reverse. Vary speed by rotating ring in either direction towards center. Note
the small depression for the current setting. After setting, the motor runs at current speed and
direction until reset. Pausing the motor is not needed to alter direction or speed using the ring.
Titan E
20,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
®
StarDental
®
Small Depression
120° Rotation Speed
Adjustment Ring
ATTACHMENTS
The Titan-E 20K Motor is designed to work with all standard E-Type attachments.
Titan E
20,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
®
StarDental
®
Straight Nose Cone
Motor-to-Angle
Contra-Angle Attachment
20K Titan-E Motor
ANGLE INSTALLATION
StarDental Titan angles can be attached to E-Type Motor-to-Angle
Contra-Angle Attachments.
To install a Titan Angle:
1. Unthread knurled ring.
2. Pull knurled ring enough to show splines, remove
knuckle from angle*.
Splines
Knuckle
Knurled Ring
Titan 3 LubeFree Ball Bearing
Auto Latch Angle Head
*
Note
: Save the knuckle in a safe place. You will
need the knuckle to use angle with Titan
attachments.
Straight Nose Cone:
Attach to motor. Install bur according to manufacturer's instructions.
NOTICE
Any serious incident occurring from the use of this device should be reported to StarDental
immediately. If incident occurred in a EU Member State the competent authority must
be notified.