User Manual
A9000
AC Power Source
Star(*)
:
Represents this command is follow the standard of IEEE 488.2, it can be used
for all instruments support IEEE 488.2 command.
Example
:
*IDN?
Explanation
:
Common command to query instrument name. Instrument will
answer its series number. DME-A902ABT A,000,000 will be the
answer for this AC source.
6.3 Command for Remote Control
6.3.1 IEEE 488.2 Standard Command
*ESE <n> Set flag mask of standard event status register.
This command is for a register used for IEE488.2 defined standard events. It is
used for flag mask or control enable. Set by 1 enables target event detection,
and event detection is masked if setting is 0. No response to target events
when 0 is set. Bits configuration for standard events register are:
bit
7
6
5
4
3
2
1
0
Bit name
PON
---- CME
EXE
DDE
QYE
----
OPC
Bit
definition
Power
on
Command
error
Execution
error
Error related
to instrument
Query
error
Operation
completed
*ESE?
Query flag mask for standard event status register.
*ESR?
Query value stored in standard events status register, 1 represents event
occurs.
*IDN?
Return string of AC source identification.
Parameter returns
:
DME-A902ABT A,000,000
DME-A902ABT A
:
Name of AC source
000,000
:
Version of firmware
*RST
Make all parameters setting back to default factory setting. Must wait 10
seconds before sending next command.
*RCL <n> Recall stored system setting.
parameter
:
1~4
*SAV <n> save system setting
parameter
:
1~4
*SRE <n> Flag mask setting for service events status register