28
2. Close the lid and securely fasten all lid clamps before proceeding. Refer to
Figure 6-3
.
Figure 6-3. Closing the lid
Filling the Chamber with Alcohol
The CUP uses a pneumatic liquid transfer pump to move the alcohol from the source tank
to the system chamber. When using a pneumatic pump, the following procedures should be
followed.
1. Verify that the hose from the source tank to the pump inlet is properly connected.
2. Verify that the hose assembly from the pump outlet to the system lid is properly
connected.
3. Set your air compressor regulator between 60 - 90 psi or applicable pump
specifications. Let the air compressor reach the desired pressure in the tank before
use.
4. Connect the air compressor throttle valve hose to the transfer pump.
5. Open the alcohol source tank valve and slowly turn the throttle valve to begin the liquid
transfer into the system.
6. Monitor the air flow to the pump to ensure the flow is consistent wit
h the manufacturer’s
recommendations for safe pump operation.
WARNING
Always visually monitor fill levels, tanks, valves, and hoses for
leakage. While the system is protected with pressure relief valves,
care should be taken to ensure overfilling does not occur.