- 23 -
5. CONFIGURATION
All transmitters have a mini-USB type, serial connector, located on the instrument
side (Fig. 5.A), through which the instrument can be connected to a PC for
configuration parameters setup.
Connection is made by means of the special RS45 (not isolated) or RS45I (isolated)
cable, which has a built-in USB adaptor and is supplied with a mini-USB connector for
the instrument plus a USB connector for the PC. For a correct connection, USB drivers
included in the CD-ROM supplied with the instrument must be installed in your PC.
Attention: the serial connection will not work if the instrument is directly connected to
a PC by means of a standard mini-USB/USB adaptor cable. For a serial connection,
you will need the special RS45 or RS45I cable.
With the RS45 cable, the instrument is powered directly by the USB port of a PC (at
least a 500 mA USB port is required), allowing in this way on-site instrument
configuration by means of a portable PC.
The PC can be connected even if the instrument has been already installed and power
supplied; the instrument is designed to avoid interferences between the two power
supplies. With the RS45I cable, the instrument is not powered by the PC USB port.
Fig. 5.A: Serial connection (mini-USB connector)
The instrument parameters setup can be easily performed by means of the supplied
application program for PC DeltaLog14 In such case, please refer to the software
manual for parameters setting.
The instrument can also be configured through the serial connection without the help
of the application program DeltaLog14, by using the serial protocol indicated in
section “COMMUNICATION PROTOCOL”.
In HD45…S… and HD46…S… models, the instrument setup can be performed
through a RS485 connection. See section “RS485 COMMUNICATION”.
HD46…DT… models can be configured by means of the front keyboard, with no need
of PC connection. For keyboard setting, please refer to section “Configuration through
keyboard”.
The instrument continues to measure normally also when it is in configuration mode;
any analog outputs or relays are regularly updated for the measured values. Changes
made to parameters have an immediate effect (from measure cycle next to parameter
confirmation).
RS45 or RS45I cable
connected to the serial port
Содержание HD45 17A
Страница 10: ... 10 3 1 HD45 connection diagrams ...
Страница 11: ... 11 3 2 HD46 connection diagrams ...
Страница 62: ... 62 Dimensions in mm HD45 series HD46 series ...
Страница 63: ... 63 Fixing holes ...