CONFIGURATION & CALIBRATION
ISO AMP
5 / 7
DELTA ELEKTRONIKA B.V.
rev. Febr. 2021
5
OPERATION & CALIBRATION
5.1
GENERAL
If you have a question about the unit, please contact our
engineers using the address [email protected].
In case the unit is defect, please first fill out the RMA-form
before sending the unit to us. Adding a detailed fault
description will help us to repair the unit as soon as
possible. Do not try to repair the unit yourself.
On our website the RMA-form can be found under
'Support'.
5.2
POWER SUPPLY SET-UP
For settings of remote programming and monitoring please
refer to the power supplies’ manual.
Note that the warning that the Ø of the programming input
of the power supply is no longer connected to the minus
DC Power Terminal, when the ISO AMP is used.
The ES- and SM-series power supplies do not need any
maintenance or calibration. However, care must be taken
that the cooling of the unit is not obstructed.
5.3
ISO AMP CONFIGURATION
The ISO AMP can be configured for both 5V and 10 V
signal levels. Selecting a signal level is done by closing or
opening DIP-jumpers.
Take proper
ESD precautions
before setting the jumpers.
See fig. 5 - 1 for lay-out for ISO AMP PCB P512 rev.B.
A closed jumper is 0- 5V, an opened jumper 0-10 V. The
signal level for each of the four analog channels can be
selected individually. See table 5 -1.
The full scale of the analog channels of the ISO AMP is
factory calibrated within 0.1% for a 5 V signal level.
Configuring jumper J1...J4 for 10 V signal levels without re-
calibrating can introduce an extra full scale error of 0.1%.
Table 5 - 2 shows the full scale calibration trimmers.
Warning:
before proceeding with the calibration of the ISO
AMP make sure that your measuring equipment has the
correct voltage and current rating.
5.4
ISO AMP VOLTAGE CALIBRATION
Connect a suitable, high accuracy volt meter to the output
terminals of the power supply and put the CV programming
switch on the power supply in the position ‘PROG’.
Apply 5.000 V or 10.000 V (depending on the jumper
setting) to the CV
PROG
input.
Calibrate the output voltage of the power supply with R51.
Connect the volt meter to the CV
MON
connection and
calibrate the CV
MON
signal with R29.
5.5
ISO AMP CURRENT CALIBRATION
Switch off the power supply, connect a suitable, high
accuracy current meter across the output terminals of the
power supply.
Put the CC programming switch on the power supply in the
position ‘PROG’.
Apply 5.000 V or 10.000 V (depending on the jumper
setting) to the CCPROG input.
Calibrate the output current of the power supply with R40.
Connect a volt meter to the CCMON connection and
calibrate the CCMON signal with R18.
Warning!
Wrong calibration can damage the unit.
5.6
Other
If there is a fault or inaccuracy that cannot be compensated
with the calibration range, please fill out the RMA-form on
our website
. See previous
paragraph 1).
fig 5 - 1
Location of jumpers and trimmers on P512 revision B.
Signal
Jumpers
CV
prg
closed=5 V, open=10 V
CC
prg
closed= +/-5 V, open= +/-10 V
CV
mon
closed=5 V, open=10 V
CC
mon
closed= +/-5 V, open= +/-10 V
table 5 -1
Jumper settings.
Signal
Trimmers (full scale)
CV
prg
R51
CC
prg
R40
CV
mon
R29
CC
mon
R18
table 5 -2
Trimmer settings.
fig 5 - 2
Legend for jumpers for ISO AMP (P512B).
fig 5 - 3
Legend for trimmers ISO AMP (P512 B).