Rooster™
Sensor200 USER MANUAL
Rooster™
Sensor200 USER MANUAL 62360MN000-A00
9 of 40
Connection and Wiring Information
Degree Controls recommends 18-24 AWG wire for all wiring connections to the input and
output connectors. The sensor and power entry connections have been prepared for you, and
merely need to be pulled through the openings and connected to the display module. If the
Sash Switch option (digital sash operation) is purchased, this is pre-connectorized for you as
well.
The Rooster
TM
Sensor200 has additional input and output capabilities for night setback and
application specific controls. These cable harnesses should be prepared by the client, using the
orderable mating connectors listed above.
Sash Input
The Rooster
TM
Sensor200 provides you with the ability to select one of two sash input settings:
1.
Digital sash sensor or
2.
Analog sash sensor
The default value is Digital, and users can toggle between a digital or analog sash sensor input
by navigating to:
Home>Settings>Sash Settings
Digital Sash
The default I/O polarity of the digital sash position input is normally open. The design intent is
as follows:
The sash setting is set to
Digital
.
When the sash is in a safe position, the proximity sensor should not detect a metallic
object/strip.
Sash position input logic may be reversed using the I/O configuration menu on the Sensor200
to achieve a different result, and any proximity sensor may be implemented.
Degree Controls can provide you with an optional Sash Switch comprised of (1) inductive
proximity sensor and (1) 10-position connector which mates to the input connection of the
display module. This inductive proximity sensor switches state when it contacts metal surfaces.
This sash sensor is a normally open device, meaning it closes upon sensing metallic objects.
The DegreeC Sash Switch (Part# 62310AS004), comes with three stripped and tinned wire leads:
Brown = +15V
Black = Sash
Blue = GND
Wire these in their respective positions 1, 2, and 3 on the input connection block.