Instruction Manual v.
1.0
Returning for Service
Returning for Service
If you need to return the
PosiTest
HHD Verifier
for service, describe the problem
fully and include reading results, if any. Be sure to include contact information
including your company name, company contact, telephone number and fax
number or email address. defelsko.com/support
The
PosiTest
HHD Verifier
is a calibrated high voltage meter used to verify
operating voltage of both Pulse DC and Continuous DC holiday detectors in the
range from 500 to 35,000 Volts.
Introduction
Introduction
Limited Warranty, Sole Remedy and Limited Liability
Limited Warranty, Sole Remedy and Limited Liability
DeFelsko's sole warranty, remedy, and liability are the express limited warranty, remedy,
and limited liability that are set forth on its website:
www.defelsko.com/terms
www.defelsko.com
© DeFelsko Corporation USA 2020. All Rights Reserved
This manual is copyrighted with all rights reserved and may not be reproduced or
transmitted, in whole or part, by any means, without written permission from DeFelsko
Corporation.
DeFelsko and PosiTest are trademarks of DeFelsko Corporation registered in the U.S. and
in other countries. Other brand or product names are trademarks or registered trademarks
of their respective holders.
Every effort has been made to ensure that the information in this manual is accurate.
DeFelsko is not responsible for printing or clerical errors.
ALWAYS ensure that the ground lead is plugged into the
PosiTest
HHD Verifier
and that the other end of the ground lead is connected to the ground
lead of the high voltage holiday detector that is being verified. Failure to ensure
proper grounding may result in electrical shock.
WARNING:
DO NOT touch the TIP of the probe or the RED High Voltage Shield
with bare hands while the probe is in contact with a high voltage source. Electrical
shock could occur if operator touches the tip or shield with bare hands while in use.
WARNING:
DO NOT use this equipment in damp, wet or rainy conditions.
Electrical shock could occur if the operator uses this equipment in damp, wet or
rainy conditions.
WARNING:
Pulse DC (30Hz) or DC
Input Impedance:
1Mohm
Max Input Power Dissipation:
12.5 watts
Range:
500–35,000 volts
Resolution:
500–1,000 volts in 10 volt increments
1,000–35,000 in 100 volt increments
Accuracy:
+/-5%
Operating Temperature Range:
-20° to 60° C (-4° to 140° F)
Battery:
3 AAA alkaline
Weight:
520 g (1.15 lbs.) with batteries and 3 m (10 ft) ground lead and clip
Dimensions:
43 x 5 x 5 cm (17” x 2” x 2”)
RED High Voltage Shield
Probe Tip
Power Button
Display
Ground (earth) wire jack
DO wear gloves and appropriate footwear
IMPORTANT:
Battery Replacement
Battery Replacement
Replace with 3 AAA alkaline batteries as shown.
Technical Data/Specifications
Technical Data/Specifications